PROBLEM SYMPTOM CHECK THIS
REMEDY
No power
Low
batteries
Error
massage
Display is dim or
will not light up.
Show on
the display
shows
Data communication
has failed
Make sure that the phone’s
Bluetooth is on or within
the distance range
E 2 shows
The cuff is very tight
Refasten the cuff and then
measure again.
E 3 shows
The pressure of the
cuff is excess.
Relax for a moment
and then measure again.
E 9 shows
Product has not been
activated.
Reactivate
E 10 or
E 11 shows
The monitor detected
motion while measuring.
Movement can affect the
measurement. Relax for a
moment and then
measure again.
E 20 shows
E 21 shows
Relax for a moment and
then measure again.
EExx,shows on
the display.
A calibration error
occurred.
Retake the measurement.
If the problem persists,
contact the retailer or our
customer service
department for further
assistance. Refer to the
warranty for contact
information and return
instructions.
The measurement
process does not detect
the pulse signal.
Loosen the clothing on
the arm and then
measure again.
Measure incorrectly.
Power is exhausted.
Charge the device
Power is low.
Charge the device
20
Power supply
3.7V 420mAH Built-in rechargeable lithium-ion
battery, 5V / 1A USB cable
Display moder
Digital LCD V.A.46.5x36.5mm
Measurement mode
Oscillographic testing mode
Measurement range
Accuracy
Normal working condition
Storage & transportation
condition
Measurement perimeter
of the wrist
About 13.5cm-21.5cm
Net Weight
Approx.110g
External dimensions
Attachment
Approx.79.8×72.5×13.2mm
USB cable,user manual
Mode of operation
Continuous operation
Degree of protection
Type BF applied part
Protection against
ingress of water
IP22, It means the device could protected
against solid foreign objects of 12.5 mm and
greater, and against vertically falling water
drops when ENCLOSURE tilted up to 15°
Software version
Device classification
V01
Battery Powered Mode:
Internally Powered ME Equipment
WARNING: No modification of this equipment is allowed.
Rated cuff pressure:
0kpa-40kpa (0mmHg-300mmHg)
pulse value:(40-199)beat/minute
Pressure:
5
℃
-40
℃
within±0.4kpa(3mmHg)
pulse value:±5%
Temperature:-20
℃
to 60
℃
RH: 10% to 93%
Atmospheric pressure: 50kPa to 106kPa
Temperature:5
℃
to 40
℃
Relative humidity ≤85%
Atmospheric pressure: 86kPa to 106kPa
AC Adaptor charged Mode:
Class II ME Equipment
21
Measurement pressure: 5.3kPa-30.7kPa
(40mmHg-230mmHg)
SPECIFICATIONS
TROUBLESHOOTING
This section includes a list of error messages and frequently asked
questions for problems you may encounter with your pressure
monitor. If the product is not operating as you think it should,
check here before arranging for service.