PROBLEM SYMPTOM CHECK THIS

REMEDY

No power

Low

batteries

Error

massage

Display is dim or
will not light up.

Show on

the display

    shows

Data communication 

has failed

Make sure that the phone’s 

Bluetooth is on or within 

the distance range

E 2 shows

The cuff is very tight

Refasten the cuff and then

measure again.

E 3 shows

The pressure of the

cuff is excess.

Relax for a moment

and then measure again.

E 9 shows

Product has not been 

activated.

Reactivate

E 10 or 

E 11 shows

The monitor detected

motion while measuring.

Movement can affect the

measurement. Relax for a 

moment and then 

measure again.

E 20 shows

E 21 shows

Relax for a moment and

then measure again.

EExx,shows on 

the display.

A calibration error

occurred.

Retake the measurement.

If the problem persists,

contact the retailer or our

customer service 

department for further 

assistance. Refer to the 

warranty for contact 

information and return 

instructions.

The measurement 

process does not detect 

the pulse signal.

Loosen the clothing on 

the arm and  then  

measure again.

Measure incorrectly.

Power is exhausted.

Charge the device

Power is low.

Charge the device

20

Power supply

3.7V 420mAH Built-in rechargeable lithium-ion 

battery, 5V / 1A USB cable

Display moder

Digital  LCD  V.A.46.5x36.5mm

Measurement mode

Oscillographic testing mode

Measurement range

Accuracy

Normal working condition

Storage & transportation

condition

Measurement perimeter

of the wrist

About 13.5cm-21.5cm

Net Weight

Approx.110g

External dimensions

Attachment

Approx.79.8×72.5×13.2mm
USB cable,user manual

Mode of operation

Continuous operation

Degree of protection

Type BF applied part

Protection against 

ingress of water

IP22, It means the device could protected

against solid foreign objects of 12.5 mm and

greater, and against vertically falling water

drops when ENCLOSURE tilted up to 15°

Software version

Device classification

V01

Battery Powered Mode: 

Internally Powered ME Equipment

WARNING: No modification of this equipment is allowed.

Rated cuff pressure: 

0kpa-40kpa (0mmHg-300mmHg)

pulse value:(40-199)beat/minute

Pressure:
5

-40

within±0.4kpa(3mmHg) 

pulse value:±5%

Temperature:-20

 to 60

℃ 

RH: 10% to 93%

Atmospheric pressure: 50kPa to 106kPa

Temperature:5

 to 40

 Relative humidity ≤85%

Atmospheric pressure: 86kPa to 106kPa

AC Adaptor charged Mode: 

Class II ME Equipment

21

Measurement pressure: 5.3kPa-30.7kPa 

(40mmHg-230mmHg) 

SPECIFICATIONS

TROUBLESHOOTING

 This section includes a list of error messages and frequently asked 

questions for problems you may encounter with your pressure 

monitor. If the product is not operating as you think it should, 

check here before arranging for service.

Содержание LS-810-B

Страница 1: ...al for further reference in case you have problems Thank you for selecting the IC Sciences Corporation Blood Pressure Monitor LS810 B Wrist Type Version 1 0 MEM SET FCC ID OU9LS810 B01 Distributed by Informed Clinical Sciences Corporation 1 Joy Street Boston USA MA 02108 Download and install My BP Coach from iTunes Store or Google play My BP Coach ...

Страница 2: ...get from the hospital different from home Is the result the same if measuring on the right wrist TROUBLESHOOTING 20 SPECIFICATIONS 21 THE MATCHED COMPONENTS 22 CONTACT INFORMATION 22 Safety Information Power Supply and Charge Power Activate Your Blood Pressure Monitor Setting Date and Time Positioning the Cuff Start Measurement Data Management 14 Recall the Records Delete the Records COMPLIED STAN...

Страница 3: ... pressure measurement Do not confuse self monitoring with self diagnosis This unit allows you to monitor your blood pressure Please start or end medical treatment basing solely on physician s treatment advice If you are taking medication consult your physician to determine the most appropriate time for your measurement Never change a prescribed medication without your physician s consent This unit...

Страница 4: ... the power Error Memory Recalling the history records Bluetooth Successful Bluetooth Connection IHB Detector Irregular Heartbeat Detector Error 5 SYSTOLIC DIASTOLIC MEM UP BUTTON SET DOWN BUTTON START STOP BUTTON CUFF Type BF applied part PULSE RATE TIME USB Interface Monitor Components List 1 Blood Pressure Monitor LS810 B 2 USB Cable 3 User Manual MEM SET Component list of pressure measuring sys...

Страница 5: ...can cause the battery to explode 4 You can only use the IC Science s authorized USB cable to charge the power You cannot use the blood pressure monitor during charging 5 During charging the blood pressure monitor displays the symbol When the charging is finished please unplug the device 6 When charging do not touch the charging connector and the patient simultaneously 7 Setting Date and Time 2 As ...

Страница 6: ...ch from the bottom of the wrist 5 Fasten the wrist cuff around your wrist leaving no extra room betweenthe cuff and your skin If the cuff is too loose the measurement will not be accurate Rest for 5 minutes before measuring your blood pressure Wait at least 3 minutes between measurements This allows your blood circulation to recover For a meaningful comparison try to measure under similar conditio...

Страница 7: ...itor and it will complete the measurement process Start Measurement CAUTION Interference may occur in the vicinity of equipment marked with the following symbol And LS810 B may interfere in the vicinity of electrical equipment Keep the monitor at least 20 centimeters 8 inches away from the human body especially the head when the data transmission is proceeding after measurement To enable the data ...

Страница 8: ...AF premature ventricular beats and peripheralarterial disease PAD should avoid using the unit whenever possible 1 When using this device please pay attention to the following situations which may interrupt blood flow and influence blood circulation of the patient thus cause harmful injury to the patient too frequent and consecutive multiple measurements the application of the CUFF and its pressuri...

Страница 9: ...ord 60 is dropped from the list CAUTION DATA MANAGEMENT 14 When you did not obtain an accurate measurement you can clear all the results by following the steps below Delete the Records 1 Under Memory Recalling Mode press and hold both the MEM button and the SET button for 3 seconds 2 The LCD will display dEL dONE indicating that memory clearing is complete Then it will shutdown To cancel the clear...

Страница 10: ...it is in proper working conditions before use Don t service or maintain while the device is in use 2 If you have any problems with this device such as setting up maintaining or using please contact Informed Clinical Sciences Do not open or repair the device yourself 3 Please report to Informed Clinical Sciences if any unexpected operation or events occur 4 Cleaning A dusty environment may affect t...

Страница 11: ...as reached a dangerous point CAUTION SYS 135 mm Hg DIA 85 mm Hg Blood Pressure Category Normal Prehypertension High Blood Pressure Hypertension Stage 1 High Blood Pressure Hypertension Stage 2 Hypertensive Crisis Emergency care needed Systolic mmHg upper Diastolic mmHg lower less than 120 120 139 140 159 160 or higher Higher than 180 and or or or or less than 80 80 89 90 99 100 or higher Higher th...

Страница 12: ...splay moder Digital LCD V A 46 5x36 5mm Measurement mode Oscillographic testing mode Measurement range Accuracy Normal working condition Storage transportation condition Measurement perimeter of the wrist About 13 5cm 21 5cm Net Weight Approx 110g External dimensions Attachment Approx 79 8 72 5 13 2mm USB cable user manual Mode of operation Continuous operation Degree of protection Type BF applied...

Страница 13: ... Symbols to be used with medical device labels labelling and information to be supplied General requirements EN 1041 2008 Medical equipment manufacturers to provide information IEC 60601 1 2005 A1 2012 Medical electrical equipment Part 1 General requirements for basic safety and essential performance IEC EN 60601 1 2 2007 Medical electrical equipment Part 1 2 General requirements for basic safety ...

Страница 14: ... This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment This transmitter must not be co located or operating in conjunction with any other antenna or transmitter EMC Guidance 24 25 1 The Blood Pressure Monitor needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the ACCOMPAN...

Отзывы: