14
Instruction Manual C100-AE
The programming of a master value is base of the Automatic Zero Adjustment.
On Automatic zero adjustment
( do this by placing the master piece into the measuring fixture and press
the
Ï
and
Ð
keys simultaneously during Measuring mode
) the master value will be taken as the measured
value and appears on the numeric display.
When measuring comparative values ( shows the deviation to the nominal size on numeric and column
displays ) zero is programmed as master value.
When pneumatic converters are used, Automatic Gauge Calibration can be considered as an option. If this
option is chosen in the Setup menu
"AutoCA"
you will be prompted to program a second master value.
If you then enter the same value for both masters, the error message
'Err. 11'
will appear. If you enter a
difference between the two masters with more than 3 mm the error message
'Err 12'
will appear.
Notice :
On Auto Zero Adjustment the zero position of the column gauge related to the master is
automatically adjusted by an offset value.
On Automatic gauge calibration the offset value ( zero point ) and the gain ( pneumatic
spread ) is calculated automatically by the column gauge.
The nominal value indicates the value to be chosen as zero point of the column display. If the nominal
value is 20 mm (
SEt.Pnt
. is set to 20 mm ) the column display will show a deviation of 0,1 mm for a work
piece measuring of 20,1 mm.
Notice :
The nominal value is used only for the zero position of the column display and does not
influence the numeric display.