Biotech-Lab User Manual

 

 

\\

  

28

www.ibpmedical.com

9 HDU-Sensors and Interfaces

9.1 HDU-CDTP  

Conductivity/Temperature Sensor

9.1.1 TECHNICAL DATA 

  

Conductivity

 

        Quadropole carbon Compound electrode.

 

 

4 measuring ranges.

 

 

Temperature compensation coefficient may be adjusted.

 

 

Range: 0...200.00 mS/cm

 

 

Cell constant: 0,54 1/cm

 

 

Accuracy:   0...199 uS/cm 0,3% ± 0.6 uS/cm

 

 

 

 

200...1999 uS/cm ± 6 uS/cm

 

 

 

 

2...11.99 mS/cm ± 0.06 mS/cm

 

 

 

 

12...19.99 mS/cm ± 0.03 mS/cm

 

 

 

 

20...200 mS/cm ± 0.6 mS/cm

Temperature

 

        Sensor integrated into the conductivity electrode

 

 

Range: 0...100 °C

 

 

Resolution: 0.01 °C

 

 

Accuracy: 25...40°C ± 0.05°C, otherwise ± 0.1°C

Calibration 

 

        PTB/NIST traceable

Communication Rate 

    All values five times per second

Degree of Protection

    P67

Power Supply 

 

        3.3 to 5 V, max. 80 mA via RS485- or USB-Interface

Environmental           

Ambient Operating Temperature Range: 0 to 50°C

Conditions 

        

Storage Temperature Range: -40 to 80°C

 

 

Operating and Storage Humidity: 85% max

 

 

relative humidity non-condensing from 0 to 50°C

Dimensions 

 

        124 x 34 mm

9.1.2 FLOW-THROUGH ADAPTER FOR CONDUCTIVITY AND TEMPERATURE MEASUREMENT

The adapter is suitable for measuring pressure, temperature and conductivity in flow-though mode. 
With immersion usage of the probe it is suitable for measuring temperature and conductivity 
concurrently. The preferred orientation of the adapter is vertical. The axial connector must point down 
(see picture). In flow-through mode the axial connector is to be used as inlet meanwhile the side 
connector is to be used as outlet. Before measurement it is necessary to shake the adapter slightly, 
so that air bubbles trapped inside can pass off through the outlet.

For immersion usage the probe must be disconnected from the flow-through.

Never touch the carbon ring electrode surfaces with your fingers or other objects. 
Do not buckle the probe cable.

Pressure
Outlet
Connector

Содержание HDC75

Страница 1: ...www ibpmedical com Biotech Lab User Manual Modular Product Family of Sensors Simulators and Testers HDU Sensor Family pH Temperature Conductivity Pressure Flow ...

Страница 2: ...e and or this manual on magnetic tape diskette or any other media except for the purpose of own use without written authorization of the IBP Medical GmbH is liable to prosecution Author Werner Pfingstmann Copyright C 2016 IBP Medical GmbH All rights reserved All brand names mentioned in this document are property of their respective owners Publisher IBP Medical GmbH Ikarusallee 15 D 30179 Hannover...

Страница 3: ... Sensor Details 24 8 HDC75 Display and Control Unit 27 8 1 Getting to know the HDC75 27 8 2 Technical Data 28 9 HDU Sensors and Interfaces 29 9 1 HDU CDTP Conductivity Temperature sensor 29 9 2 HDU PRH HDU PRS Pressure Sensor Family 30 9 3 HDU FL Flow Sensors 32 9 4 HDU Pt100 I Interface for Pt100 temperature sensors 33 9 5 Pt100 Sensors Temperature Sensor Family Air Surface General Dip Laboratory...

Страница 4: ...alid calibration certificates Only use the Biotech Lab system in a dry environment and do not touch it with wet hands Prevent every mechanical overstraining of electrical wires If the acquired values seem to be not believable make sure that the no device is defective Prevent electrostatic discharge on the connectors This can lead to substantial damage of the Biotech Lab devices and devices under t...

Страница 5: ...ation and use of the Biotech Lab System The system includes Display and Control Units Sensors Interface Modules Software 1 1 INTENDED USE The Biotech Lab system is intended to be used in biomedical laboratory and environmental applications Do not use any part of the system in life support applications ...

Страница 6: ...ver Products HDC75 Display and Control Unit HDU CDTP Conductivity Temperature Sensor HDU PRH High Accuracy Pressure Sensor Family HDU PRS Pressure Sensor Family HDU FL Flow Sensor Family HDU Pt100 I Interface for Pt100 temperature sensors HDU pH I Interface for pH Sensor HDU pHPt I Interface for pH Sensor and Pt100 temperature sensors Relevant EC Directives 2004 108 EC 2006 95 EC Compliance With E...

Страница 7: ...ice PC or Tablet with Microsoft Windows Depending on the requirements you can use the HDC75 or a standard computer with Microsoft Windows as display and control unit for the Biotech Lab modules Biotech Lab software works with Microsoft Windows 7 and 10 Connectivity of Biotech Lab modules All Biotech Lab modules such as sensors simulators analyser and tester feature an RS485 Interface With a specia...

Страница 8: ...nterface for Pt100 temperature sensors Available versions depending on the requirements Temperature Sensor Family Air Surface General Dip HDU YSI400 I Interface for YSI 400 temperature sensors HDU pH I Interface for pH Sensor 3 1 2 HSI SIMULATORS HDS Sources and HVS Simulators are a product family voltage and pressure sources and Vital Signs Simulators for biomedical laboratory and environmental a...

Страница 9: ... Biotech Lab server is the central software module Additional App s provide the functionality needed for the different Biotech Lab modules The App s are independent programs which can be used simultaneously All general functions like settings are also available in all different App s The server functionality includes Scanning for connected Biotech Lab Modules Communication with all Biotech Lab Mod...

Страница 10: ... file in C Users accountname IBP Medical system directory 4 3 INSTALLATION PC For PC s is a special version of the Biotech Lab available Download the software and start setup exe program The different Biotech Lab App s must be installed separately 4 4 SOFTWARE UPDATE For software updates please navigate to http www ibpmt com btle shtm Just download the Apps you want to update and copy it into rela...

Страница 11: ... 10 please navigate to http windows microsoft com en us windows 5 1 CHANGING THE DISPLAY LANGUAGE IN WINDOWS 10 To change the default language sign in to Windows 10 and Go to Start Menu and select Settings Select Time Language Select Region Language Select the region you are in Click Add a language ...

Страница 12: ...d You ll see three options Set as default Options Remove Under Options click Download to download the language pack and keyboard for that language If you open Control Panel Languages you ll see the traditional settings which you are familiar with Changing the Display Language To change the language of the user account you are currently using select a language then click Set as default ...

Страница 13: ...App allows displaying measuring values from sensors simulators and testers If you start the HDC75 the Lockscreen will look as follows Slide from bottom to top to get the Windows Start Screen Access to online manual Quick Reference and Biotech Lab support Starts the Biometer Software ...

Страница 14: ...d devices and shows the devices suitable for this program During the scan all visibla sensor values are Settings leads to the base settings of BioTechLab Downloads the user manual 6 2 HEAD AREA FUNCTIONALITY Button Function Button shown once messages are available Shows Biotech Lab and App Messages Deletes all messages Shows list of all values ...

Страница 15: ...lue display L acc is the accumulated amount in L This is a typical two value display with an additional functionality in the headline Dialysis Machine Standard indicates here Nthe selected conductivity temperature compensation coefficient Shows settings for the particular device For sensors a single sensor graphic is shown In case a sensor is disconnected the system scanns the sensor ports automat...

Страница 16: ...e following functions are available with connected mouse only With a Right Mouse Click into the graphic you get the following menu options More sub functions below leads to Graphic Settings deletes the measured values Additional Subfunction The following functions are available with connected mouse only With a Right Mouse Click into the graphic you get the following menu options More sub functions...

Страница 17: ...le by Mouse To get the following sub functions make a Left Mouse Click followed by a Right Mouse Click Scale Range allows modifying the scales on both sides of the graphic Allows selecting the time interval Allows selecting the colors of both the grid and the backplane ...

Страница 18: ...LUES After pressing Minimum Value Maximum Value Arithmetic Mean Value Standard Deviation Current Value Selection of Statistic Interval This is the time range for the calculation of the statistic values The time range starts from the last measuring value and goes back with the selected value Element Function ...

Страница 19: ...ed to 25 C In other words the display is always converted to a solution temperature of 25 C The temperature coefficient which the displayed value is compensated with is expressed as C Unfortunately however different solutions also have different temperature coefficients To achieve an exact reading the instrument will have to be adjusted to the temperature coefficient of the current solution The co...

Страница 20: ...t with the machine manual If you have machines from one manufacturer only in your use the temperature coefficient that the dialysis machine is using for compensation If you have different types of dialysis machines in your unit the best solution is to use a temperature coefficient of 2 07 C for all machines This avoids confusion with different readings of the conductivity on different machines For...

Страница 21: ...te Conductivity 2 200 K 2 200 K STANDARD Other Dialysis Machines 2 000 K 2 070 K USER TC Adjustable 0 5 000 K 0 5 000 K nLF ISO Non Linear Function for According Standard tap purified and waste water ISO7888 DIN EN27888 Reference temperature 25 C RO DWA Reverse Osmosis DWA 0 00 K RO DWA DWA GmbH Co KG Ubstadter Straße 28 76698 Ubstadt Weiher RO HERCO Reverse Osmosis Herco 2 200 K RO HERCO Herco Wa...

Страница 22: ...E SENSOR DETAILS The following menu shows the settings for pressure sensors Head area shows the serial number and the date of last calibration Select here the unit you like to see Allows selecting a higher resolution for the reading Allows taring the sensor Allow selecting the damping factor Starts a pressure change measurement 7 3 1 OFFSET BALANCING TARE Caused by drift of the pressure probe the ...

Страница 23: ...ing the damping factor 7 4 1 FLOW DIRECTION To avoid malfunction of the sensor please make sure to connect fluid correctly the sensor The sensor has a label showing the flowdirection 7 5 PH SENSOR DETAILS The following menu shows the settings for pH sensors Head area shows the serial number and the date of last calibration Invokes the adjustment function for the sensor See below Allows selecting t...

Страница 24: ...ent points Current reading of pH and temperature Allows selecting and inputting the buffer solution value Leads to the next calibration point if any Otherwise the next window shows the calibration result Once all adjustment points are finished the adjustment values are store automatically in the sensor If something is wrong with the adjustment the sensor will not accept the values and an error mes...

Страница 25: ...d by leaking electrolyte can be removed easily by rinsing with reverse osmosis treated water The sample volume should be 100 ml of dialysate or 1 liter of untreated water The sample should be poured into a clean glass container with a tube or hose coming into as little contact with air as possible The pH value must be measured immediately in the same container Make sure that the display stabilizes...

Страница 26: ...75 while in proper fast charge mode results in the sensor ports being inaccessible as long as charging is in progress To activate the sensor ports again it is necessary to briefly unplug the USB charger and then plug it in again The charge circuit will revert to normal or slow charge depending on the capabilities of the USB charger LED GreenCharger connected LED Red on Charging with a standard cha...

Страница 27: ... 3 MP fixed focus front facing 2 MP fixed focus rear facing Audio Mono speaker Interface Micro USB B Host also charger connector 5 x RS485 M8 Connector for Biotech Lab Modules Degree of Protection IP54 Power Supply 3 7 V 14000 mHh 52 Wh Li ion polymer battery Quick Charge 2 0 compatible Environmental Conditions Operating Temperature Range 0 to 50 C Storage Temperature Range 40 to 80 C Operating an...

Страница 28: ...o 50 C Conditions Storage Temperature Range 40 to 80 C Operating and Storage Humidity 85 max relative humidity non condensing from 0 to 50 C Dimensions 124 x 34 mm 9 1 2 FLOW THROUGH ADAPTER FOR CONDUCTIVITY AND TEMPERATURE MEASUREMENT The adapter is suitable for measuring pressure temperature and conductivity in flow though mode With immersion usage of the probe it is suitable for measuring tempe...

Страница 29: ...ily 9 2 1 Technical Data All Versions Reading Readings as mmHg kPa Bar or PSI Measuring Temperature compensated Pressure Connection IBP Sensor Port and G1 8 Female internal thread Membrane HDU PRS versions have a Stainless steel membrane which is fluid resistant HDU PRH5 Range 5 to 5 psi 0 34 to 0 34 bar 258 to 258 mmHg Resolution 0 01 mmHg Accuracy General 0 05 full scale Media Compatibility Gass...

Страница 30: ...edia Compatibility Gasses and Fluids Over pressure 1 5 x full scale HDU PRH50 Range 13 to 50 psi 0 93 to 3 5 bar 700 to 2580 mmHg Resolution 0 1 mmHg Accuracy 0 05 full scale Media Compatibility Gasses Over pressure 2 x full scale HDU PRH100 Range 13 to 100 psi 0 93 to 7 bar 700 to 5150 mmHg Resolution 0 1 mmHg Accuracy 0 05 full scale Media Compatibility Gasses Over pressure 2 x full scale HDU PR...

Страница 31: ...age Temperature Range 40 to 80 C Operating and Storage Humidity 85 max relative humidity non condensing from 0 to 50 C Dimensions 80 x 34 mm 9 2 2 MEASURING ADAPTER Each HDU Pressure sensor comes with tubing set which allows connecting the sensor to typical Biomedical applications To adapt the sensor to specific measuring point different adapter are available If you have special needs please do no...

Страница 32: ...easured value Repeatability better 0 5 Fluid temperature 20 C to 80 C Pressure 0 1 to 5 bar at 22 C Calibration PTB NIST traceable Communication rate Five times per second Degree of protection IP67 Power supply 3 3 to 5 V 60 mA via RS485 or USB Interface Environmental conditions Operating Temperature Range 0 to 50 C Storage Temperature Range 40 to 80 C Operating and Storage Humidity 85 max relativ...

Страница 33: ... 70 mA via RS485 or USB Interface Environmental Conditions Ambient Operating Temperature Range 0 to 50 C Storage Temperature Range 40 to 80 C Operating and Storage Humidity 85 max relative humidity non condensing from 0 to 50 C Dimensions 64 x 45 x 25 mm 9 5 PT100 SENSORS Temperature Sensor Family Air Surface General Dip Laboratory Rugged Pt100 temperature sensor family Air General Dip Surface Lab...

Страница 34: ... 0 to 50 C Dimensions 64 x 45 x 25 mm 9 7 HDU PHPT I Interface for pH Sensor and Pt100 Temperature Sensors pH Range 0 14 pH Resolution 0 01 pH Accuracy 0 02 pH Temperature Pt100 Range 50 100 C Resolution 0 001 C Accuracy 25 40 C 0 015 C otherwise 0 025 C if Sensor and interface are adjusted together Communication Rate Five times per second Degree of Protection IP67 Power Supply 3 3 to 5 V max 80 m...

Страница 35: ... US 9 8 HDU YSI400 I Interface for YSI 400 temperature sensors 9 9 HDU YSI700 I Interface for YSI 700 temperature sensors 10 HSI Simulators These modules are under development If you have special needs please do not hesitate to contact us 11 HAT Tester These modules are under development If you have special needs please do not hesitate to contact us ...

Страница 36: ...t it should be not more than one third and preferably one tenth of the permissible error of the confirmed equipment when in use In other words the accuracy of the references to adjust or calibrate the HDM must have at least a tree times higher accuracy After a channel has been adjusted it is necessary to calibrate it afterwards If inconclusive or incomprehensible values are shown in the display it...

Страница 37: ...rd solution and calibration verification accuracy Keep solutions tightly capped to avoid evaporation Do not return used solutions to the storage bottle Do not remove solutions from their original bottle Keep the solutions in a cool place Use only fresh reference solutions for calibration and verification Use the solution immediately after pouring evaporation will cause errors Discard solution the ...

Страница 38: ... clean any device with any fluids In case of pollution you can wipe the surface of the Biotech Lab devices with a dry and clean cloth 13 3 CALIBRATION PERIOD When delivered new from the factory each device is adjusted and calibrated before shipment To ensure reliability we recommend the following intervals for calibration Parameter Calibration Period Temperature Once per year Conductivity Once per...

Отзывы: