ibidi Heating System, Universal Fit
5
Operation
Before starting an experiment, check that the ibidi Heating System, Universal Fit fits on your micro-
scope stage, and that your cell culture vessels are compatible with the Heating Insert. Connect all
parts (Section
) and perform a temperature adjustment as explained in Section
NOTE
The Temperature Controller only measures the temperature from the sensors in the Heated Lid
and Heated Plate. The temperature in the sample must be adjusted for your specific setup. Follow
the instructions in Section
5.1
Installation and Connection of the Parts
The components of the ibidi Heating System, Universal Fit are connected as shown in Figure
Figure 11 – Installation of the components of the ibidi Heating System, Universal Fit.
NOTE
Before shipment, all controllers are run through an in–house calibration with the correspond-
ing Heated Lid and Heated Plate. Only use the Temperature Controller with the corresponding
Heated Lid and Heated Plate to ensure correct sensor calibration.
1. Place the Temperature Controller next to the microscope and connect the power supply.
2. Confirm that the power switch of the Temperature Controller is off.
Version 2.2 (May 23, 2019)
22
Instruction Manual