ibaPADU-C-8AI 

Manual 

 

 

Issue 1.11 

19

 

 

6.3 

Start measurement process 

First connect the input signals with the analog or digital inputs. The measurement can 
be started in two ways: 

6.3.1 

Start measurement process manually 

The manual measurement  process is started and stopped by briefly pressing the S11 
key. Proceed as follows: 

1. 

Switch the device on by briefly pressing the S11 key. All LEDs initially light up in red, 
once the device is ready L1 lights up orange. 

2. 

Start the measurement by briefly pressing the S11 key. As long as the measurement 
process is running, the L1 will flash in green. 

3. 

You can stop the measurement by briefly pressing the S11 key again. L1 lights up 
orange. 

6.3.2 

Start measurement process triggered 

If the measurement process should be started via a trigger signal, then it must be entered 
in the configuration file "config.ini": 

Configuration for "trigger": [highPulse] 

The  measurement  process  only  takes  as  long  as  the  trigger  signal  is  pending  (digital 
signal, start trigger increasing flange, stop trigger decreasing flange). The trigger signal 
must be pending at least as long as the configured sample rate in order to start the data 
recording.  

The trigger signal connection is described in chapter 5.4.4 

“Analog/digital inputs X1”. 

Proceed as follows: 

1. 

Switch the device on by pressing the S11 key for a longer period. All LEDs initially 
light up in red, once the device is ready L1 lights up orange.  

2. 

Go to the "wait for trigger" state by briefly pressing the S11 key. L1 lights up green. 

3. 

L1  flashes  green  when  the  trigger  signal  is  pending  and  the  measurement  starts. 
Data is only saved while the trigger signal is pending. 

4. 

Once the trigger signal is no longer pending, the measurement is ended and L1 lights 
up green. 

5. 

To leave the "waiting for trigger" state, briefly press S11. 

 

 

Note 

During each measurement a new measurement file is created. During long-term record-
ings new measurement files are automatically created after every 100,000 samples. 

Once the free memory space on the internal memory falls below 40 MB, the measure-
ment data recording is automatically aborted. This is indicated by the LED L1 flashing 
red. In addition, an error message file "Error.log" is stored in the "log" directory with the 
following content: „[ERROR]    stopped measuring --> running out of disk space“. 

Содержание ibaPADU-C-8AI

Страница 1: ...ibaPADU C 8AI Mains Independent Data Logger Manual Issue 1 11...

Страница 2: ...d trademark of the Microsoft Corporation Other product and company names mentioned in this manual may be labels or registered trademarks of the corre sponding owners Certification The device is certif...

Страница 3: ...mory 13 6 Commissioning and configuration 14 6 1 Requirements 14 6 2 Configuration 14 6 2 1 Time synchronization 17 6 2 2 Product information 18 6 3 Start measurement process 19 6 3 1 Start measuremen...

Страница 4: ...capable of assessing safety and recognizing possible consequences and risks on the basis of his her specialist train ing knowledge and experience and knowledge of the standard regulations 1 2 Designat...

Страница 5: ...ess and the system or machine to be controlled there is a risk of death or severe injury The non observance of this safety information may result in a potential risk of death or severe injury The non...

Страница 6: ...onnected to a PC via an USB connection The PC recognizes the devices per plug and play like a mass storage device In addition it is possible to retrieve the data using the network connectivity via FTP...

Страница 7: ...lable Strain relief plate Manual 4 Safety instructions Maintaining the operational voltage range Operate the device only with a voltage of 9 30 V DC Too high of an operational voltage will destroy the...

Страница 8: ...C 8AI 8 Issue 1 11 5 Description of the device 5 1 Device overview 1 Operational state display LEDs L1 L4 2 S11 key 3 DC input X14 4 Ethernet interface X22 5 USB interface X23 6 Analog digital inputs...

Страница 9: ...es orange Lights up red Battery is almost empty Flashes red Error status All LEDs light up red Boot process All LEDs flash orange Update is being loaded All LEDs are off Not in operation no power supp...

Страница 10: ...not connected as a mass storage via USB and an IP address has been configured 5 4 3 USB interface X23 A PC can be connected via this USB interface in order to read data At the same time the internal b...

Страница 11: ...the analog digital inputs E 1 2 E 10 V A0 3 4 A4 10 V GND 5 6 GND A1 7 8 A5 GND 9 10 GND A2 11 12 A6 GND 13 14 GND A3 15 16 A7 GND 17 18 GND GND 19 20 GND D0 21 22 D4 D1 23 24 D5 D2 25 26 D6 D3 27 28...

Страница 12: ...is identical to the digital input design There are 2 ways to con nect the trigger signal Signal level against GND Signal level Trigger 2 2 V active 0 9 V signal level 2 2 V undefined 0 9 V inactive ex...

Страница 13: ...d with the following directory structure Directory Contents app ibaAnalyzer application config Configuration file config ini Example of configuration file template_config ini dat Directory of the meas...

Страница 14: ...hernet The following parameter can be set sampling rate trigger time zone hours and minutes data format dat or csv DHCP device IP address subnet mask gateway IP address for time synchronization signal...

Страница 15: ...5 Possible settings dhcp This setting determines whether the IP address of the device should be assigned by a DHCP server in the network ipAddress IP address of the device This setting will be ignored...

Страница 16: ...ted by the auto cleanup function paduc_2014 12 31_23 59 59 dat Only empty directories with the following format are deleted 2014 12 31 name Names can be assigned to the 8 analog and 8 digital input si...

Страница 17: ...log directory which indicates the current error in clear text 6 2 1 Time synchronization The time can be synchronized via an external server The timeserver address is stored in the device default time...

Страница 18: ...lable measurement will be started without synchronization 6 2 2 Product information The file product log is stored on the device in the log directory It contains important device information such as h...

Страница 19: ...e trigger signal must be pending at least as long as the configured sample rate in order to start the data recording The trigger signal connection is described in chapter 5 4 4 Analog digital inputs X...

Страница 20: ...ose the explorer and remove the device symbol from the system bar with Securely remove hardware Note The measurement data should be deleted after successfully transmitting to the PC in order to always...

Страница 21: ...sists of the product name and the 6 digit serial number of the device You find the serial number on the type label on the device Example serial number 000001 Hostname paduc_000001 The hostname can als...

Страница 22: ...ctory 4 Close the explorer and remove the device symbol from the system bar 5 Restart the device by pressing the S11 key longer than 3 s All LEDs then extinguish and after a brief period the device re...

Страница 23: ...as residual waste For proper disposal or exchange send the complete device via post using sufficient stamps back to iba The batteries are manufactured with components made of valuable materials which...

Страница 24: ...ge Digital inputs Number 8 Version Single ended no galvanic isolation Input signal range 0 V to 30 V Signal level log 0 0 9 V Signal level log 1 2 2 V Sampling rate Linked with analog sampling Communi...

Страница 25: ...Operational and ambient conditions Operational temperature range 32 F to 122 F 0 C to 50 C Storage temperature range 4 F to 140 F 20 C to 60 C Transport temperature range 4 F to 140 F 20 C to 60 C Hum...

Страница 26: ...onnection plug arrangement of the analog digital inputs PE 1 2 PE A0 3 4 A4 GND 5 6 GND A1 7 8 A5 GND 9 10 GND A2 11 12 A6 GND 13 14 GND A3 15 16 A7 GND 17 18 GND GND 19 20 GND D0 21 22 D4 D1 23 24 D5...

Страница 27: ...al number iba S N of the product Contact Headquarters iba AG Koenigswarterstr 44 90762 Fuerth Germany Phone 49 911 97282 0 Fax 49 911 97282 33 Email iba iba ag com Contact Mr Harald Opel Shipping addr...

Отзывы: