SW-4000T-MCL User Manual (2.0)
Setting List
- 65 -
AcquisitionControl
Configure image capture settings.
Acquisition Control Item Setting Range
Default Value
Description
AcquisitionLineRate (Hz)
Related Topic:
66 to 67702
(steps of 0.1)
66
Set the AcquisitionLineRate(Hz).
TriggerMode
Related Topic:
0:Off
1:On
0:Off
Select the trigger mode.
TriggerSource
-
20: Line4 TTL
In1
Select the trigger signal source.
0: Low
1: High
10-13: PulseGenerator0-3
14-17: UserOutput0-3
20: Line4 TTL In1
21: Line5 Opt In1
23: Line7 CC1
24: Line10 TTL In2
26: NAND0 Out
27: NAND1 Out
28: Line13 TTL In3
29: EncoderTrigger
TriggerActivation
0:Rising Edge
1:Falling Edge
3:Level High
4:Level Low
1:Falling Edge
(falling edge of
input signal)
Select the polarity of the trigger signal (i.e., location of signal at
which trigger is applied).
ExposureMode
Related Topic:
0: Off
1: Timed
2: TriggerWidth
1: Timed
Select the exposure mode.
ExposureTimeMode
0: Common
1: Individual
0: Common
Select the Exposure setting method.
Select
Common
to set a common value for all three sensors.
Select
Individual
to set the value for each channel.
Note:
This item is available only when
ExposureMode
is set
to
Timed
.