IACER SRL
MNPG05-07 Ed. 29/03/10 Pag. 53
A09 • Distortion/Contusion (Sprains / Bruises)
(Medical program)
(only on
T-ONE EVO II)
The program develops its effectiveness after this type of injury by inhibiting pain locally,
producing three selectively acting, differentiated impulses. The intensity should be adjusted to
a level between the thresholds of perception and pain:
Number of treatments: until pain is lessened, on a daily basis (even 2/3 times a day).
A8 Sprains/Bruises
Total duration: 30 minutes
Electrodes Positioning:
ankle (photo 32).
Phase 1
Phase 2
Tens rapid 15 min.
Haematomas impulse 15 min.
A10 • Inflammatory
(Medical program)
(only on T-ONE EVO II)
Program recommended for inflammatory conditions. To be applied until the inflammatory state
is lessened (10-15 applications, once a day; the daily treatments can be doubled if required).
Identify the area to be treated and position the electrodes as shown in illustration 1. Adjust the
intensity until a tingling feeling is produced in the area treated; avoid contracting the
surrounding muscles.
Program duration: 30 minutes.
A11 • Vascularization
(Medical program)
(only on T-ONE EVO II)
Has a vascularizing effect on the area treated. Vascularization increases arterial flow and
consequently aids the removal of algogenic substances and helps to restore normal
physiological conditions. Do not position the electrodes close to inflamed areas.
Daily application is recommended, the number of applications is not defined; the program can
be used to reduce pain.
Stimulation intensity should be between the thresholds of perception and slight discomfort.
Program duration: 20 minutes.
Position of electrodes: see photo 25 and 33 in the manual of positions.
A12 • Decontracting
(only on T-ONE EVO II)
Can be used for all sports, after competitions or the most demanding training sessions, in
particular after long and intense exertion. To be used immediately after exertion. Helps
drainage and capillarization, improving muscle oxygenation and helping to discharge synthetic
substances produced during exertion. Program duration: 35 minutes. Stimulation intensity:
medium-low, increased during the last 10 minutes.
Position of electrodes: see photos 01 to 20 in the manual of positions.
A12 Decontracting
Total duration: 35 minutes
Electrodes Positioning:
form a square above the painful area as shown in illustration 1.
Fase 1
Fase 2
Tens a impulso stretto 15 min.
Tens a impulso largo 20 min.
A13 • Hematoma
(Medical program)
(only on T-ONE EVO II)
Consult a doctor before using this program to treat haematomas. Total program duration: 20
minutes. Few applications carried out within a few hours of the bruise. A combination of
various types of square-wave impulses has a graduated draining effect on the area to be
treated (impulses at different frequencies drain the area at different depths). The intensity
should be adjusted to a level between the thresholds of perception and pain, without causing
muscle contraction.
Position of electrodes: form a square above the area to be treated as shown in illustration 1.
Содержание T-ONE EVO I