IACER Srl
48
MNPG148-06
and carefully stored on a secure surface. The performances of the equipment
are granted if it is stored according to the following conditions:
Outside the carrying bag:
Temperature
from 5 to +40°C
Relative humidity
from 30 to 75%
Pressure
from 700 to 1060 hPa
Inside the supplied carrying bag:
Temperature
from -10 to +55°C
Relative humidity
from 10 to 90%
Pressure
from 700 to 1060 hPa
N.B.
Disconnect the cables, before storing the device into its carrying bag. If
not, the cables could bend excessively near the connectors. It could severely
damage the cables.
Troubleshooting
Any type of work on MIO-CARE PRO must be carried out exclusively by the
manufacturer or by an authorized dealer. In any event, any presumed
malfunction of MIO-CARE PRO must be verified before sending the device to
the manufacturer.
Here below are some typical situations:
•
MIO-CARE PRO cannot be turned on and/or the display does not light
up:
-
Check the battery status and replace it if it is necessary (make
reference to chapter
Battery replacement
).
If the problem persists contact the manufacturer
•
MIO-CARE PRO does not transmit electric impulses
:
−
Check that the cable jacks have been inserted in the electrodes and
that the plastic protection has been removed from the electrode.
−
Check that the cables have been connected correctly (connector well
inserted in the device).
−
Check that the cables and the electrodes are not damaged.
If the problem persists contact the manufacturer.
•
MIO-CARE PRO transmits low intensity or intermittent impulses
:
−
Check the cables and the electrodes are in good condition and
replace them if it is necessary.
If the problem persists contact the manufacturer.
•
MIO-CARE PRO switches off during the operation:
−
It is suggested to replace the battery and start a new treatment.
If the problem persists contact the manufacturer.