MNPG148-06
Prg
Medical
prg
Yes/No
Description
PHASE 1
PHASE 2
PHASE 3
Frequency 6
Hz
impulse
width 250
µ
s
250
µ
s x 8sec
CH1&CH2
80%+ 50Hz
250
µ
s x 6 sec
75% CH1 +
50Hz 250
µ
s x
6 sec 75%
CH2) x 45
cycles
Frequency 10
Hz
impulse width
250
µ
s (80%)
15
No
Sequential
tonic
contractions –
upper limbs
and trunk
Total time 3
min
Frequency 6
Hz
impulse
width 200
µ
s
Total time 10
min (30Hz-
200
µ
s x 5 sec
80% CH1 +
30Hz-200
µ
s x
5 sec 80%
CH2) x 60
cycles
Total time 5
min
Frequency 4
Hz
impulse width
200
µ
s (90%)
16
No
Sequential
tonic
contractions –
lower limbs
Total time 3
min
Frequency 6
Hz
impulse
width 300
µ
s
Total time 10
min (30Hz-
300
µ
s x 5 sec
80% CH1 +
30Hz-300
µ
s x
5 sec 80%
CH2) x 60
cycles
Total time 5
min
Frequency 4
Hz
impulse width
300
µ
s (90%)
17
No
Sequential
phasic
contractions –
upper limbs
and trunk
Total time 3
min
Frequency 6
Hz
impulse
width 200
µ
s
Total time 10
min (50Hz-
200
µ
s x 5 sec
75% CH1 +
50Hz-200
µ
s x
5 sec 75%
CH2) x 60
cycles
Total time 5
min
Frequency 4
Hz
impulse width
200
µ
s (90%)
18
No
Sequential
phasic
contractions –
lower limbs
Total time 3
min
Frequency 6
Hz
impulse
width 300
µ
s
Total time 10
min (50Hz-
300
µ
s x 5 sec
75% CH1 +
50Hz-300
µ
s x
5 sec 75%
CH2) x 60
cycles
Total time 5
min
Frequency 4
Hz
impulse width
300
µ
s (90%)