How to look
after the device
Checking device operation
A magnet (small ring or disc in metal or metal/plastic) is supplied with the
appliance to check device operation.
Procedure for checking:
1. switch on the device following all the safety instructions provided in
this manual;
2. start any therapy, following the instructions for use of this manual;
3. hold the supplied magnet and bring it closer to the applicator;
4. check that the magnet vibrates (proportional to the frequency of the
selected therapy).
Contact the manufacturer if the magnet fails to vibrate.
Cleaning the device
Use a soft dry cloth to remove any dust from the device.
More difficult stains can be removed using a sponge soaked in a water
and alcohol solution (20% alcohol).
To clean the 3-solenoid belt or the circular cases of the professional
applicator with a pair of solenoids, it is recommended to disconnect the
applicator from the device before carrying out any operation.
•
Remove the 3-solenoid cable by removing the 2 silver studs with
a screwdriver or open the circular cases using the side zip.
•
Clean the fabric with water and neutral soap and wait for it to dry
completely before replacing the applicators. The fabric used is
designed to withstand 10 cleaning cycles while still maintaining its
original characteristics.
CAUTION: always respect the polarity of the applicators taking care to
insert the coils with the side indicated by the + sign towards the green part
of the belt (therapeutic side).