Intended purpose and scope of use 

Clinical purpose: 

 

 

Therapeutic 

Scope of use: 

   

Outpatient Clinic/Hospital and home 

 

LaMagneto  is  designed  and  indicated  for  the  treatment,  rehabilitation 
and functional recovery of conditions concerning: 

 

 

joints of the wrist, hand, shoulder, foot, ankle and knee 

 

the musculoskeletal system 

 

arthrosis    

 

muscular atrophies and dystrophies                         

 

bursitis 

 

bruises      

 

degenerative disease of the locomotor system 

 

sprains   

 

periarthritis                  

 

benign lesions and pulled muscles 

 

tendinitis and tendinosis  

LaMagneto  is  particularly  indicated  for  the  treatment  of delayed  union, 
osteoporosis,  bone  oedema,  osteonecrosis,  as  well  as  ulcers  and 
neuropathies. 

Thanks to the high intensity of the magnetic field it  is able to generate, 
LaMagneto  is  particularly  indicated  in  the  treatment  of  bone  fractures 
even in the presence of rigid bandages or plaster casts.  

The  distinctive  feature  of  the  LaMagneto  device  is  the  possibility  of 
managing  two  treatments  at  the  same  time,  by  setting  different  times 
and programs. 

LaMagneto  is  a  device  intended  for  both  professional  users  (doctor, 
therapist etc.) and for in-home therapy. With regard in-home therapy, 
it  is  recommended  that  the  device  only  be  used  following  advice 
from your doctor/therapist. 

In  accordance  with  guidelines  for  medical  devices,  the  manufacturer 
suggests  a  check  of  the  efficiency  and  safety  of  the  device  every  24 
months. Useful life of the device and its accessories (period after which 
it is suggested to send the device to the manufacturer):3 years 

 

Содержание LaMagneto

Страница 1: ...USER MANUAL MNPG262 02 Edition27 01 2020 Magnetotherapy model LaMagneto I A C E R Srl www itechmedicaldivision com ...

Страница 2: ...I A C E R Srl Via S Pertini 24 A 30030 Martellago VE ITALY Tel 39 041 5401356 Fax 39 041 5402684 e mail iacer iacer it https www itechmedicaldivision com ...

Страница 3: ...edevice 11 Warnings 11 Electromagneticinterference 12 Contraindicationsandsideeffects 13 Quickuseofthedevicewithpresetparameters 13 Useofthetherapeuticbeltsandsolenoids mainapplicationsandsuggestions 15 Instructionsforusingthepresetprograms 17 InstructionsforusingprogrammesinDoublePatientmode 20 Listofstoredprograms 23 Last10 28 Settings languageselection 29 Howtolookafterthedevice 30 Checkingdevi...

Страница 4: ...sed as the most suitable technique for the treatment of bone conditions and in particular for osteoporosis The biological modifications induced by the magnetic fields on the cell membranes guarantee a biostimulation able to restore the correct functionality of the cell itself According to the experiences of several authors in cases of osteoporosis already starting from the sixth treatment session ...

Страница 5: ...plies with the essential requirements of Council Directive 93 42 EEC of 14 June 1993 concerning medical devices Leg Decree46 97 of 24 February 1997 Implementation of Directive 93 42 EEC concerning medical devices Annex II as amended by Directive 2007 47 EC of 5 September 2007 Leg Decree37 2010 of 25 January 2010 Notified body TÜV Rheinland Italia srl Via Mattei 3 20010 Pogliano Milanese MI Italy C...

Страница 6: ...d magnetic fields do not fall into the category of ionising radiation and that do not involve direct application of electric current on the patient it can be said that the energy exchanged by the device with the human body is not at all dangerous Therefore the devices of the LaMagneto line are active medical devices belonging to class IIa With regard to the second sentence of Annex IX If several r...

Страница 7: ...europathies Thanks to the high intensity of the magnetic field it is able to generate LaMagneto is particularly indicated in the treatment of bone fractures even in the presence of rigid bandages or plaster casts The distinctive feature of the LaMagneto device is the possibility of managing two treatments at the same time by setting different times and programs LaMagneto is a device intended for b...

Страница 8: ...ield Adjustable with increasing scale up to 150 Gauss per channel Frequency of the square wave 1 100 Hz Therapy time User settable The maximum magnetic field intensity is 150 Gauss per channel with a professional applicator with a pair of solenoids optional accessory The values of intensity frequency and time are supplied with an accuracy of 20 Ambient operating conditions temperature from 5 to 30...

Страница 9: ...I A C E R Srl 9 MNPG262 02 Label 1 on the back Labelling Devicelabelsindetail Label 1 Label 2 Label 3 Label 4 DC 15V 2 0A ON OFF CH1 CH2 UES36LCP 150200SPA Label 2 Label 3 Label 4 ...

Страница 10: ...ean Community Directive 93 42 EEC and subsequent amendments Date of manufacture month year Serial number Temperatures permitted Relative humidity Manufacturer s data IP21 Degree of protection against the entry of solids powders and liquids Center positive symbol Contentsofthepack The LaMagneto pack contains N 1 LaMagneto device N 1 medical power supply approx 1 5mt cable N 1 Use and maintenance ma...

Страница 11: ...ess please follow the following instructions Check the location and meaning of all labels affixed to the device Do not damage the applicator by acting on the connecting wire also avoid winding the wire around the applicator or around the device Check the integrity of the power supply each time it is used Avoid use in the case of signs of damage to the casing or to the connecting wire People who ar...

Страница 12: ...connect the power supply from the wall socket at the end of the therapy session The manufacturer is to be considered responsible for the safety reliability and performance of the device provided that any additions modifications and or repairs are carried out by personnel authorised directly by the manufacturer the electrical system of the environment in which LaMagneto is inserted complies with na...

Страница 13: ...electromagnetic field generated by the device Quickuseofthedevicewithpresetparameters To start using LaMagneto quickly and easily we recommend that you follow the steps below 1 Connect the applicator or applicators to the device by connecting the plug of the applicator cable to one of the two ports CH1 CH2 on the panel at the top of the appliance 2 Connect the mains cable to the power supply and t...

Страница 14: ...ht below the display notifies the therapy is underway 8 At the end of the therapy the device will automatically return to the program menu screen Note it is possible to temporarily suspend therapy at any time by pressing the OK button To resume therapy press the OK button again During the pause phase the green LED goes off and then comes back on when the therapy is restarted Note it is possible to...

Страница 15: ...itions for applying the therapeutic belt and the solenoids are given below N B it is recommended to interpose the non woven fabric strip between the skin and the applicator Wrap the 3 solenoid belt applicator around the area to be treated or place it over the area such as in the treatment of the vertebral column taking care that the green side of the belt is placed in contact with the skin Two pho...

Страница 16: ...I A C E R Srl 16 MNPG262 02 Ankle Hip Back lumbar Knee Shoulder Wrist Elbow Cervical vertebrae Knee Tibia fibula Ankle Shoulder Femoral head Femur ...

Страница 17: ...enoids available as an optional accessory for LaMagneto Instructionsforusingthepresetprograms Per utilizzare LaMagneto impostando liberamente i parametri relativi a tempo di terapia e intensità del campo magnetico seguire i seguenti semplici passi 1 Connect the applicator or applicators to the device by connecting the plug of the applicator cable to one of the two ports CH1 CH2 on the panel at the...

Страница 18: ...ttons and select the desired program 6 Press OK The display will show the basic setting of therapy time 2 hours and intensity of magnetic field that we will modify a Press the button so that the spanner is highlighted At this point press OK the display shows a moving spanner icon on the left hand side ...

Страница 19: ...py hours from 0 to 24 and confirm by pressing the OK button The display will highlight the minutes of therapy c Use the and buttons to set the desired therapy minutes from 0 to 59 and confirm by pressing the OK button The display will highlight the intensity of the treatment ...

Страница 20: ...aMagneto device follow these simple steps 1 Connect the applicator or applicators to the device by connecting the plug of the applicator cable to one of the two ports CH1 CH2 on the panel at the top of the appliance 2 Connect the mains cable to the power supply and then connect the power supply outlet to the circular connector on the panel at the top of the device near the ON OFF switch 3 Connect ...

Страница 21: ... settings a Use the and buttons to set the desired therapy hours from 0 to 24 and confirm by pressing the OK button The display will highlight the minutes of therapy b Use the buttons and to set the desired therapy minutes from 0 to 59 and confirm by pressing the OK button The display will highlight the intensity of the treatment ...

Страница 22: ...the therapy is restarted Note it is possible to exit the treatment at any time by pressing the button once the device will go back to the screen of the selected programme point 6 By pressing the button again the device will go back to the initial screen of the programme menu point 5 Note the device recognises if the applicators are connected correctly During the therapy phase the connection status...

Страница 23: ...n 30 2 6 30 60 24 hours 6 Cervicalgias 20 2 6 30 60 24 hours 7 Chronic pain 10 2 6 30 60 24 hours 8 Fractures 25 2 6 30 60 24 hours 9 Epicondylitis 25 2 6 30 60 24 hours 10 Epitrocleitis 25 2 6 30 60 24 hours 11 Pseudoarthrosis 75 2 6 30 60 24 hours 12 Lumbalgy 50 2 6 30 60 24 hours 13 Lumbar pain 50 2 6 30 60 24 hours 14 Shoulder Arthrosis 30 2 6 30 60 24 hours 15 Knee arthrosis 30 2 6 30 60 24 h...

Страница 24: ... 24 hours 31 Discopathy 25 2 6 30 60 24 hours 32 Myalgia 1 2 6 30 60 24 hours 33 Neuropathy 10 2 6 30 60 24 hours 34 Muscle strain 1 2 6 30 60 24 hours 35 Muscular cramp 1 2 6 30 60 24 hours 36 Rhizarthrosis 25 2 6 30 60 24 hours 37 Impingement syndrome 50 2 6 30 60 24 hours 38 Carpal tunnel 50 2 6 30 60 24 hours 39 Titanium prosthesis 75 2 6 30 60 24 hours 40 Rotator cuff 73 2 6 30 60 24 hours 41...

Страница 25: ... specific program for the stimulation of bone regeneration in a post traumatic condition Epicondylitis specific program for this condition aimed at promoting the recovery of the joint in an inflammatory condition of the tendons Epitrocleitis specific program for this condition aimed at promoting the recovery of the joint in an inflammatory condition of the tendons Pseudoarthrosis specific program ...

Страница 26: ...designed for the treatment of inflammation of tendon tissues Chondropathy specific program for osteoarthritis a particular type of chondropathy designed for the reduction of inflammation of cartilaginous tissues Anti inflammatory program to mitigate inflammatory and painful conditions Useful also in cases of post surgery recovery Whiplash syndrome program designed for post traumatic recovery with ...

Страница 27: ...onal recovery and pain reduction for this type of condition Tarsal tunnel syndrome program designed to reduce the pain of this condition and other polyneuropathies Programs from 42 to 45 night vary in frequency but are the same in terms of waveform They are indicated for a therapy time ranging from 6 to 10 hours during sleep After long and exhaustive bibliographic research the correspondence betwe...

Страница 28: ...1974 Bassett Valdes Hernandez 1982 Last10 This mode allows direct access to the last 10 therapies used by the user From the main menu press until Last 10 is selected then press OK Choose the therapy from those listed by pressing the and buttons followed by OK Choose the desired settings a Use the and buttons to set the desired therapy hours from 0 to 24 and confirm by pressing the OK button The di...

Страница 29: ...OFF switch located on the small panel in the upper part so it is in the ON position After the I TECH MEDICAL DIVISION logo appears press and select the Settings menu At this point select Language and use the and buttons to select the desired language N B to turn off the device press the ON OFF switch on the back or press the button until the screen turns off ...

Страница 30: ...he device More difficult stains can be removed using a sponge soaked in a water and alcohol solution 20 alcohol To clean the 3 solenoid belt or the circular cases of the professional applicator with a pair of solenoids it is recommended to disconnect the applicator from the device before carrying out any operation Remove the 3 solenoid cable by removing the 2 silver studs with a screwdriver or ope...

Страница 31: ...ave the following characteristics ambient temperature from 5 to 40 C relative humidity from 15 to 93 pressure from 700 to 1060 hPa Disposal The product is subject to the WEEE regulation the symbol is present on the label concerning separate collection to dispose of the product make use of special areas equipped to collect electronic material by contacting the competent authorities in your country ...

Страница 32: ...ntact for technical support regarding the device For all technical support matters please contact I A C E R S r l Via S Pertini 24 a 30030 Martellago VE Tel 39 0415401356 Fax 39 0415402684 Technical documentation concerning repairable parts may be provided but only with prior company authorisation and only after giving proper training to the maintenance personnel Spareparts Original spare parts fo...

Страница 33: ...ddition to domestic ones and those directly connected to a low voltage power supply network that supplies buildings for domestic use The device can be used in all buildings including domestic buildings and those directly connected to the public low voltage power supply network that supplies buildings for domestic use Harmonic emissions IEC 61000 3 2 Class A Complies The device can be used in all b...

Страница 34: ...ironment Impulses EN 61000 4 5 1kV differential mode 1kV differential mode Mains power quality should be that of a typical business or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines EN 61000 4 11 0 UT for 0 5 cycles different angles 0 UT for 1 cycle 70 UT for 25 30 cycles 0 UT for 5 seconds 0 UT for 0 5 cycles different angles 0 UT for 1 cy...

Страница 35: ...istance in meters m Radiated RF EN 61000 4 3 10 V m 80 MHz to 2 7 GHz 10 V m 80 MHz to 2 7 GHz The field strength from fixed RF transmitters as determined by an electromagnetic site survey may be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol Recommended separation distance between portable and mobile radi...

Страница 36: ...uation applicable to the frequency of the transmitter where P is the maximum output power of the transmitter in watts W according to the transmitter manufacturer Note 1 At 80 MHz and 800 MHz the highest frequency range applies 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people ...

Страница 37: ...er with all accessories To be eligible for warranty work the purchase must send the appliance with the receipt or invoice proving the correct origin of the product and the date of purchase Warranty conditions 1 Should assistance be needed enclose the purchasing receipt when sending the device to the manufacturer 2 The warranty period 2 years is valid only on the electronic parts The warranty will ...

Страница 38: ...I A C E R Srl 38 MNPG262 02 LaMagneto All rights reserved LaMagneto and the logo are the exclusive property of I A C E R Srl and are registered trademarks ...

Страница 39: ......

Страница 40: ......

Отзывы: