26
MNPG118-02
degassed (by previous boiling) in order to prevent air bubbles that could
decrease the effectiveness of the treatment.
WARNING
. Do not apply directly on the head of the handle the gel: it
could emit the ultrasound beam, therefore ruining the head itself.
Always use a CE marked gel.
WARNING.
In order to guarantee the patient safety, the device stops
emitting the ultrasound beam and alerts the users with the blinking
LED when the head exceeds the 42°C. The device will emit again when
the temperature drops under 41°C.
WARNING. Only in the ultrasound mode the handpiece can be used
for immersion treatments. The handpiece and its cable are the only
parts protected against water damage with an IPX7 level.
Programs features and main applications
Refer to the following table for programs features. All parameters are
adjustable by the user.
PROG.
Medical
prg.
YES/NO
FREQ.
DUTY
CYCLE
TIME
SUGGESTED
INTENSITY
P-01
Yes
1MHz
50%
14min
1.0W/cm²
P-02
Yes
1MHz
50%
20min
1.0W/cm²
P-03
Yes
1MHz
50%
20min
1.0W/cm²
P-04
Yes
1MHz
50%
30min
0.5W/cm²
1.0W/cm²
P-05
Yes
3MHz
50%
16min
1.0W/cm²
P-06
Yes
1MHz
50%
14min
1.0W/cm²
P-07
Yes
1MHz
50%
14min
1.0W/cm²
P-08
Yes
1MHz
50%
14min
1.0W/cm²
P-09
Yes
1MHz
50%
14min
1.0W/cm²
P-10
Yes
3MHz
50%
14min
1.0W/cm²