3-13
Calibrations and Quality Assurance
032-0328-EN Rev K
11. Record the Mean HU value of each material. See table below. Recorded values should
fall within the lower and upper limits.
12. When finished, save the file and close TxSTUDIO.
Material
Mean Scan Value
(Hounsfield Units)
Lower
Limit
Upper
Limit
Mean
Air
(Black) (lower left)
-1000
-980
-990
Acrylic
(Light Gray) (lower right)
-50
200
75
LDPE
(Dark Gray) (upper right)
-250
-50
-150
Teflon
(White) (upper left)
580
1160
870