background image

 

HyQual Multi-probes User Manual v3, 08.07.2022 
KISTERS Controlled, Template: Word-with-table-of-contents-Template-Company.docx, v1.1 

Page 15 / 23 

 

Turbidity 

2 points 

0 NTU, 10 NTU, 100 NTU, 400 NTU 

calibrate  near  expected 
value 

HDO (Optical DO) 

calibrate  at  100%  saturated 
water 

DI  water  -shake  vigorously  to 
oxygenate 

set  BP  before  calibrating, 
recal  at  deployment  site 
for best accuracy 

Chlorophyll 

2 points  

secondary  solid  or  40µg/L  solution 
or lab sample 

 

Rhodamine 

2 points 

secondary 

solid 

standard 

or 

rhodamine 

 

Blue Green Algae 

2 points 

secondary  solid  standard  or  lab 
sample 

 

Ammonium (NH4+) 

2 points 

Lo 4.63 mg/l; Hi 46.3 mg/l 

 

Nitrate (NO3+) 

2 points 

Lo 4.62 mg/l; Hi 46.2 mg/l 

 

Chloride (CL-) 

2 points 

CD Standard 147 Micro S  
CD Standard 1412 Micro S 

enter 34.3 mg/l for low cal  
enter 319.3 mg/l for high cal 

7.3.4

 

Calibration Record (Cal Log) 

Every HyQual probe has a dedicated data file called CAL.LOG. The CAL.LOG records every calibration 
made to your instrument, whether you accepted the calibration or not. This file shows the time and date 
of the calibration, the parameter calibrated, the reading before the calibration was accepted, the read-

ing after the calibration was accepted, the “raw” sensor reading, the SRF, and a few other details.  If you 

wish to know, for instance, the last time that DO was calibrated, the Calibration Record would tell you 
when the most recent DO calibration was made, the value of the calibration standard, and the instru-

ment’s reading in the standard before the calibration was made (to tell you exactly how much the in-

strument was changed during calibration). Please notice that this data cannot be altered within the 
HyQual probe!. 

 

7.3.5

 

Temperature 

The Temperature sensor is an electrical resistor (thermistor) whose resistance changes predictably with 
temperature. The sensor is protected by a stainless-steel tube. Thermistors are very stable with time, so 
are factory-calibrated and do not require recalibration. 

7.3.6

 

Dissolved Oxygen 

The optical dissolved-oxygen sensor is a blue-light source, a sensing surface, and a red-light receiver.  
The sensing surface is an oxygen-active compound stabilized in an oxygen-permeable polymer, usually 
silicone.  When the sensing surface is exposed to oxygen (in water or air), oxygen diffuses into the sens-

Содержание HyQual

Страница 1: ...HyQual Water Quality Multi Probe HyQuest Solutions KISTERS Group 2022 05 03...

Страница 2: ...ction 10 6 2 2 Unattended Logging 10 6 2 3 Telemetry Relay 11 6 2 4 On Line Monitoring 12 7 1 When to proceed to maintenance 12 7 2 Maintenance 1 Keep your device clean 12 7 3 Maintenance 1 Calibratio...

Страница 3: ...ilidad de tales da os Cualquier error encontrado en cualquier producto de HyQuest Solutions debe ser comuni cado a HyQuest Solutions donde se har todo lo posible para resolver r pidamente el pro blema...

Страница 4: ...industrial specification that defines the electrical interface and physical layer for point to point communication of electrical devices The RS 485 stand ard allows for long cabling distances in elec...

Страница 5: ...s battery operates only at the surface and its job is to power the multiprobe while you re pairing the instrument with a data display via Bluetooth wireless technology A USB port on a PC is another wa...

Страница 6: ...send the probe back to us at your own expense Additional sensors Barometric pressure is calculated with the use of the depth sensor therefore you will need to add an optional depth sensor to any of yo...

Страница 7: ...ufacturer s named repair center You can find information on how to re turn items for repair in the relevant section of the HyQuest Solutions website Disposal instructions After taking any HyQuest Solu...

Страница 8: ...ufficient water flow that can be easily ac cessed for maintenance and protected from external elements such as animals and items such as stones or sticks that may damage the probe 4 1 Protection for u...

Страница 9: ...n to toggle between Logging ON and Logging OFF The Setup function for Logging lets you name your Logging file and set the logging interval time be tween logged lines of data Figure 1 Switch Logging ON...

Страница 10: ...The Display can be a laptop or almost any tablet or smart phone You can record measurements using the HyQual probe s Snapshot or Automatic Snapshot features You might snapshot a series of measurements...

Страница 11: ...rchase this battery after as it is not integrated to the probe Please note that this battery ads several inches to the length of the probe and that it can also be used with the HyQual 200 only that it...

Страница 12: ...ed in the control loop or decision making process Maintenance 7 1 When to proceed to maintenance Judgment gained from observing your field conditions and data requirements provides information about w...

Страница 13: ...tuation for which the correct parameter value is known Here s the general procedure 1 Clean the sensor and perform any necessary sensor specific maintenance 2 Select a calibration standard whose value...

Страница 14: ...c But you can elect to accept any SRF 7 3 3 Choosing Calibration Standards For best results choose a calibration standard whose value is close to what you expect to see in the field For example calibr...

Страница 15: ...ate of the calibration the parameter calibrated the reading before the calibration was accepted the read ing after the calibration was accepted the raw sensor reading the SRF and a few other details I...

Страница 16: ...follow the calibration instructions on the screen The manufacturers of optical DO sensors usually recommend that you not calibrate the zero DO point because the zero DO point drift is very low However...

Страница 17: ...don t report because it has little meaning in natural waters You can t calibrate TDS or salinity directly because they are calculated from Conductivity You can however calibrate TDS with a TDS standar...

Страница 18: ...ous circle at the end of the sleeve This not only purges bubbles from the electrolyte but also cleans nasty stuff out of the junction Figure 4 Filling in fresh pH reference electrolyte 7 3 10 ORP ORP...

Страница 19: ...s a small hole in the back of the transducer There is a tube inside the cable i e a Vented Cable to connect the Stage sensor to the atmosphere so that changes in barometric pressure will not affect t...

Страница 20: ...e small set screw on the wiper arm 2 Remove the wiper pad from the wiper arm and replace the pad 3 Place a new wiper arm on the motor shaft so that the set screw faces the flat spot on the motor shaft...

Страница 21: ...bor charge 1 2 days down time Repair HyQuest Solutions precision instruments and data loggers are produced in quality controlled processes All HyQuest Solutions production and assembly sites in Austra...

Страница 22: ...lay devices not for eld use calibration solutions copper anti fouling kit USB adapter connect DB9 to USB inte grated SDI 12 and MODBUS output SDI 12 adapter cable MODBUS adapter cable and Android disp...

Страница 23: ...ion battery must be dis posed of according to legal provisions EU WEEE Directive As players in the environmental market KISTERS AG and HyQuest Solutions are committed to supporting efforts to avoid a...

Отзывы: