background image

 

HyQual Multi-probes User Manual v3, 08.07.2022 
KISTERS Controlled, Template: Word-with-table-of-contents-Template-Company.docx, v1.1 

Page 13 / 23 

 

 

Figure 3 

 Getting access to O-Rings 

Always remove batteries (if any) and clean The HyQual probe prior to storing it for prolonged periods. 
Always refill the reference electrode and recalibrate after long storage periods. 

7.3

 

Maintenance 

 1: Calibration 

7.3.1

 

Basics of Parameter Calibration 

The procedure for operating a HyQual, including making calibrations, varies with the type of Display 
used because the user-interface software can be different (mostly because of the differences in Display 
screen sizes).  You could also study the highly detailed Appendices to familiarize yourself with the vari-
ous procedures.  Otherwise, you should be able to walk through the software once you have established 
communication between your Display and the HyQual probe. You can become a minor expert in just a 
few minutes. 
The HyQual probe never guesses parameter values, so you must calibrate it from time to time by simply 
telling the instrument what it should read in a calibration situation for which the correct parameter 

value is known.  Here’s the general procedure:

 

1.

 

Clean the sensor and perform any necessary sensor-specific maintenance. 

2.

 

Select a calibration standard whose value is close to the values you expect to see in the field.  
For best results, use fresh calibration solutions, and discard them once they have been used.  
But generally, you can reuse most calibration standards a few times if you are careful to avoid 
contamination. 

3.

 

With the HyQual probe

’s Storage/Calibration Cup screwed onto 

the HyQual probe housing, rinse 

the sensors three times with a small quantity of your calibration standard by pouring the stand-

ard into the Storage/Calibration Cup, positioning the “stopper” side of the lid (the side with the 

O-ring) on top of the Storage/Calibration Cup, and shaking the HyQual probe vigorously to re-
move traces of old calibration solutions.  Discard the used calibration standard between rinses.   

4.

 

Next secure The HyQual probe with the sensors pointing up and fill the calibration cup with your 
calibration standard.  Make sure the standard covers the sensor entirely, and that it also covers 
the thermistor for those parameters that are temperature-compensated. For turbidity sensors 

and other fluorometers fill the cup to at least 1 ½ inches above the sensor’s lens surface.

 

Содержание HyQual

Страница 1: ...HyQual Water Quality Multi Probe HyQuest Solutions KISTERS Group 2022 05 03...

Страница 2: ...ction 10 6 2 2 Unattended Logging 10 6 2 3 Telemetry Relay 11 6 2 4 On Line Monitoring 12 7 1 When to proceed to maintenance 12 7 2 Maintenance 1 Keep your device clean 12 7 3 Maintenance 1 Calibratio...

Страница 3: ...ilidad de tales da os Cualquier error encontrado en cualquier producto de HyQuest Solutions debe ser comuni cado a HyQuest Solutions donde se har todo lo posible para resolver r pidamente el pro blema...

Страница 4: ...industrial specification that defines the electrical interface and physical layer for point to point communication of electrical devices The RS 485 stand ard allows for long cabling distances in elec...

Страница 5: ...s battery operates only at the surface and its job is to power the multiprobe while you re pairing the instrument with a data display via Bluetooth wireless technology A USB port on a PC is another wa...

Страница 6: ...send the probe back to us at your own expense Additional sensors Barometric pressure is calculated with the use of the depth sensor therefore you will need to add an optional depth sensor to any of yo...

Страница 7: ...ufacturer s named repair center You can find information on how to re turn items for repair in the relevant section of the HyQuest Solutions website Disposal instructions After taking any HyQuest Solu...

Страница 8: ...ufficient water flow that can be easily ac cessed for maintenance and protected from external elements such as animals and items such as stones or sticks that may damage the probe 4 1 Protection for u...

Страница 9: ...n to toggle between Logging ON and Logging OFF The Setup function for Logging lets you name your Logging file and set the logging interval time be tween logged lines of data Figure 1 Switch Logging ON...

Страница 10: ...The Display can be a laptop or almost any tablet or smart phone You can record measurements using the HyQual probe s Snapshot or Automatic Snapshot features You might snapshot a series of measurements...

Страница 11: ...rchase this battery after as it is not integrated to the probe Please note that this battery ads several inches to the length of the probe and that it can also be used with the HyQual 200 only that it...

Страница 12: ...ed in the control loop or decision making process Maintenance 7 1 When to proceed to maintenance Judgment gained from observing your field conditions and data requirements provides information about w...

Страница 13: ...tuation for which the correct parameter value is known Here s the general procedure 1 Clean the sensor and perform any necessary sensor specific maintenance 2 Select a calibration standard whose value...

Страница 14: ...c But you can elect to accept any SRF 7 3 3 Choosing Calibration Standards For best results choose a calibration standard whose value is close to what you expect to see in the field For example calibr...

Страница 15: ...ate of the calibration the parameter calibrated the reading before the calibration was accepted the read ing after the calibration was accepted the raw sensor reading the SRF and a few other details I...

Страница 16: ...follow the calibration instructions on the screen The manufacturers of optical DO sensors usually recommend that you not calibrate the zero DO point because the zero DO point drift is very low However...

Страница 17: ...don t report because it has little meaning in natural waters You can t calibrate TDS or salinity directly because they are calculated from Conductivity You can however calibrate TDS with a TDS standar...

Страница 18: ...ous circle at the end of the sleeve This not only purges bubbles from the electrolyte but also cleans nasty stuff out of the junction Figure 4 Filling in fresh pH reference electrolyte 7 3 10 ORP ORP...

Страница 19: ...s a small hole in the back of the transducer There is a tube inside the cable i e a Vented Cable to connect the Stage sensor to the atmosphere so that changes in barometric pressure will not affect t...

Страница 20: ...e small set screw on the wiper arm 2 Remove the wiper pad from the wiper arm and replace the pad 3 Place a new wiper arm on the motor shaft so that the set screw faces the flat spot on the motor shaft...

Страница 21: ...bor charge 1 2 days down time Repair HyQuest Solutions precision instruments and data loggers are produced in quality controlled processes All HyQuest Solutions production and assembly sites in Austra...

Страница 22: ...lay devices not for eld use calibration solutions copper anti fouling kit USB adapter connect DB9 to USB inte grated SDI 12 and MODBUS output SDI 12 adapter cable MODBUS adapter cable and Android disp...

Страница 23: ...ion battery must be dis posed of according to legal provisions EU WEEE Directive As players in the environmental market KISTERS AG and HyQuest Solutions are committed to supporting efforts to avoid a...

Отзывы: