6
3.2.4 Waste and reagent tubing
Connect supplied waste tubing to the waste outlet.
Washing solution
Temperature
control unit
Washing solution
Reagent 1
Reagent 2
Reagent 3
The Innovate System Hardware User Guide
www.hygiena.com
3.2.3 Reagent tray
Assemble the reagent tray using the Allen key supplied with the Innovate system.
The tray contains two tube holders, one with an opening and one without an
opening. Place the tube holder with the opening at the front side of the tray.
The liquid level will be visible through the opening. Place the cooling unit
inside the tray.
• Add a few drops of antifoam to the waste container.
• Place waste tubing into the waste container to collect the waste solution.
• Instrument is shipped with external reagent tubing connected to
injectors 1, 2, and 3.
• Place all three injector needles in washing solution.
Note: Wear gloves when touching the injector needles.