19
Error Messages - Causes and Solutions
14
Err
or Messages - Causes and Solutions
Display: ERR
The device could not read the information from the RFID card.
Elimination
Press the button briefly (<1 second), the display will show
ON
and continue
with point 5.1 or 5.2. If the error occurs several times, please contact your
supplier.
Display: DATE
The expiry date of the test seems to be exceeded.
Elimination
The device checks the internally set time with the expiration date of the test.
Check the expiration date of the test. If this is actually exceeded, select a
new test from another shipment which has not expired. Briefly press the
button. The display will show
ON
and repeat point 4a or 4b. If the expiration
date is not exceeded, check the internally set time and date of the device
and adjust it if necessary.