Hydro-Air P-38
Installation
pH / ORP Analyzer
7
Conductive layer not removed
pH
Ref.
Conductive layer removed
pH
Ref.
4.2.
Sensor Connection
Extension of sensor cable
If the user chooses to extend the sensor cable, please ensure the extension cable is
coaxial type and is in good condition.
For sensors without a built-in pre-amplifier (e.g. Conventional Combination Sensor), it is
highly recommended to install an external pre-amplifier into any cable connection
exceeding 25m to ensure excellent reading accuracy and consistency of the sensor.
Removal of Conductive Layer
The user must ensure the pH cable is separated from the Reference cable. Moreover,
the conductive layer covering the pH cable must be removed in order to isolate the pH
circuit connection from the Reference. Otherwise, the sensor will display incorrect
readings.