4. Mounting the Electrodes:
If applicable, install the necessary hardware to mount the electrode. This
may be a separate acrylic pot or a PVC gland that threads into the fl ow cell.
1. Remove the electrode from its buff er cap and place it into its mounting gland. Secure it into place
by gently threading it into the gland, rotating clockwise until it stop.
2. Connect the quick-disconnect cap & cable assembly to the top of the pH or ORP electrode by
gently threading the cap on, rotating clockwise until it stops.
5. Wiring the Electrodes:
The WQM-100 will have its electrode connections pre-wired and only the
electrode(s) need to be connected. If an electrode is being added that the unit was not originally
supplied with or if the cable assembly is being replaced, please referece the following installation
points.
NOTE: If installing an electrode that the equipment was not supplied with, it may be necessary to
add an additional liquid tight cable grip into the controller enclosure. This hardware will have been
provided with the new electrode that is to be installed.
1. With the instrument powered off , run the cable portion of the electrodes quick-disconnect cap &
cable assembly through the liquid tight cable grip in the controllers enclosure.
2. Refer the circuit board fi gure (FIgure 8) for wiring connection terminals for the electrode that is
being connected.
6. Installation Inspection:
Reference the troubleshooting table in this document if needed.
•
Make sure that the sensing glass is fully submerged.
•
Make sure that the sensing glass is free of air bubbles.
•
There should be a reading on the instrument display.
FIGURE 4 (Mounting Gland, Quick-Disconnect Cap & Cable)
WARNING! Electrical hazards are involved with this installation. Only qualifi ed personnel
should perform this installation.
7