HydrINS 2 / HydrINS 2 Mini flow meter
User Manual
Page 43 / 79
8.5.5
External power supply by military connector
For connection to the military connector, use a CNT104/CNT105 cable, and then connect the banana
plugs to an external power supply.
1 – Remove all connections from Connector J15, and then connect the additional board, Cat. No.,
4005511 to the same connector. In this case, the position of Switch SW2 used earlier is no longer
relevant. Switches SW1 and SW2 of the additional board will be used.
2 - Create the following configuration:
Switch SW1 of the additional board:
MilSpec
Switch SW2 of the additional board:
External
8.6
Electrical power supply of Display C
Risk of electrocution. Always disconnect the power supply before making any electrical
connections.
Potential damage to the device. The internal electronic components of the device can
be damaged by static electricity, which might affect its performance characteristics and
operation.
Display C may be powered in two different ways:
Terminal: By external power supply connected to the 20-36 VDC (20-28 VDC for serial
numbers below 41574) terminals,
MilSpec: By power supply via the military connector and a CNT 104 or CNT 105 cable.
Select the appropriate switch position indicated on the image below: