5.2
Method for installation at 1/8
Figure 16: Diagram of installation at 1/8 diameter
For an installation at 1/8 diameter, the distance Z
1/8
after probe insertion
must be equal to:
Z
1/8
= L - E - 1/8 DI
Z
1/8
: Distance between the lower edge of the guide bar and the
top of the pipe (mm or in)
E: Pipe thickness (mm or in)
D: Inside diameter of pipe (mm or in)
L: Length (mm or in) – value according to model, equal to the
distance between the lower edge of the guide bar and the two
measurement electrodes located on the sensor
Model
L mm (inches)
300
760 (29.92)
500
935 (36.81)
700
1175 (46.26)
1000
1475 (58.07)
Mini
575 (22.64)
Table 2: Correspondence of lengths L with the HydrINS probe model for an installation at
1/8th diameter
Z
1/8
E
1/8 DI
DI
Guide Bar