Using the switching outputs
MCS1xxx /-T005
en(us)
Page 39 /
68
BeWa MCS1xxx-T005 4668567a en-us 2022-06-14.docx
2022-06-14
You can set the duration of the particle signal between 5 - 200 Ms.
This value is factory set to 7 ms.
The signal gives no indication of the shape and size of the particles. The
duration of alarm signal is set to 30 seconds
Note that after the sensor has been switched on or after voltage
has returned, a start-
up phase of ≈
70 seconds is necessary
before the first measurement can be made.
Example: Operation - 1 particle detected = logical "active"
Example: Operation - no particle detected = logical "not active"
Example: Operation - several particles detected = logical "active"