29 / 51
1.7.0 Volume table
For calibrating the instrument for volume measurement.
1.7.2. Input table
Select point 01 to 20, enter level and then
volume values respectively.
This function is used for setting up the strapping table (level/volume). Up to 20 points
can be assigned. Every new point must be larger than the preceding one. The units of
length and volume can be changed later without affecting the settings in the table.
The units selected for length and volume in Functions 1.2.4 and 1.2.5 will be used here.
0
(conversion table not created, volume
measurement not possible)
5.1.3
C
ONFIGURATION
:
CONFIGURATION EXAMPLES
The minimum functions to be configured for a simple measurement are listed below:
Functions
Definition
HyView
HART®
1.1.1
2.1.1.1
Tank Height
1.1.2
2.1.1.4
Dead zone
1.5.1
2.1.5.3
Detection delay
1.3.1 … 4
2.1.3.1 … 4
Current Output
1.7.0
2.1.7.0
Volume table*
*For volume measurements
Example procedures are given on the following pages. Each procedure is given in a series of steps in table form and is according to the HyView Configuration
parameter list. Please refer to section 5.3.4 for the equivalent parameters available on the HART® Communicator.
Definitions for configuration
Where HyView user menu Function 1.3.1 (HART® Function 2.1.3.1) Current I1 Item is configured to “Level”.
Typical instrument used for quick configuration examples:
Probe type:
twin Ø4 mm or 0.15” cable probe
Product measured:
Water (dielectric constant,
r
= 80)
Tank height
10000 mm
Dead zone
(see “ Technical data of the probes”)
Probe length L
2
9000 mm (do not modify unless advised to)