2.1 Platinum wire radius: the radius of the platinum ring wire. Normally it is 0.30 mm.
Click the data box to pop up the numeric keyboard, enter the data.
2.2 Platinum ring radius: the radius of the platinum ring. Normally it is 9.8 mm. Click
the data box to pop up the numeric keyboard, enter the data.
2.3 Density of heavy liquid: The density of the sample with heavy density (for
example, water is 1.000). Click the data box to pop up the numeric keyboard, enter
the data.
2.4 Light liquid density: the density of light samples (for example, transformer oil is
0.875). Click the data box to pop up the numeric keyboard, enter the data.
2.5 After setting the parameters, press the "Save button" to save the current
parameters. Press the "X" button to return to the previous menu. (All data has been
entered and confirmed when leaving the factory, and there is no need to modify it if
there are no special circumstances.)
3. Pure water calibration
3.1 Preparations (implementingArticle 4.1 of BG6541) Clean the sample cup with
petroleum ether, then wash with acetone and water respectively, and then wash with
hot chromic acid lotion to remove oil stains, and finally rinse with water and distilled
water. If the sample is not used immediately, it should be poured into a clean On the
cloth. Clean the platinum ring with petroleum ether, rinse with methyl ethyl ketone,
and heat it in the oxidizing flame of an alcohol lamp. Note: To clean the platinum ring,
it should be removed from the rod of the ring frame. When removing and installing
the platinum ring, be sure to handle it gently.After installing the platinum ring, make
sure that each part of the platinum ring is on the same plane (Figure 5). ). Hang the
cleaned platinum ring. Pour pure water into the clean sample cup to the middle scale