126
126
HPE-810
HUVITZ
8.
System Maintenance
8.1.
Automatic Calibration of Frame Reader
Automatic Calibration of Frame Reader is required with the following cases,
When the Size or Axis value of the traced data seems to have a problem,
When the Frame Reader seems to have a problem in operating normally.
Automatic calibration of frame reader consists of three parts: Stroke, Frame, Pattern. If any changes
in Stylus occur, Stroke calibration must be executed at first. The order of calibration is like this:
Stroke -> Frame -> Pattern.
8.1.1.
Stroke Calibration
Procedures for stroke calibration
①
Press MENU button and select CALIB.
②
Press FRAME button and select STROKE.
③
Press BOTH button.
④
Pull the Gripping bars to the maximum.
⑤
Press BOTH button to start calibration.
Next, follow the instructions displayed.
8.1.2.
Frame Calibration
Procedures for frame calibration
①
Press MENU button and select CALIB.
②
Press FRAME button and select FRAME.
③
Press BOTH button.
④
Load the Standard Frame.
⑤
Press BOTH button to start calibration.
Next, follow the instructions displayed.
Содержание HPE-810
Страница 1: ...1 1 User Manual EDGER HPE 810...
Страница 19: ...19 19 Devices for Edging Polycarbonate...
Страница 87: ...87 87 Mirroring the slot Mirroring the group of hole slot...
Страница 123: ...123 123 How to adjust rear safety bevel value Add value How to adjust front safety bevel value Add value...
Страница 131: ...131 131 8 3 2 X Y R You can check X Y R axis X axis test Y axis test R axis tes X Y R sensor state 3 4 1 2...
Страница 135: ...135 135 8 3 6 Clamp Window You can check clamp and window Clamp test Window test Clamp and window sensor state 3 1 2...
Страница 153: ...153 153 10 3 Drawings of System 10 3 1 Drawings of Edger...
Страница 154: ...154 154 HPE 810 HUVITZ 10 3 2 Drawings of Frame Reader 10 3 3 Drawings of Blocker...