With the UV, we can calibrate the UV value of our machine.
The procedures are as followings:
A.
Press the ‘UV’ (
) button.
B.
Open UV cover.
C.
Focus to UV 0.0 by using ‘INDEX’ button.
D.
‘-0.01’button and ‘+0.01’ button by the given value in the UV lens specification.
E.
When UV transmissivity is 0.0
: Prevent part that UV-LED comes out.
When UV transmissivity is 25.0/50.0/75.0
: Put a lens on the top of the UV-LED comes out.
When UV transmissivity is 100.0
: Nothing raises.
F.
Press the ‘MEASURE’ button.
G.
UV transmissivity by 0.0->25.0->50.0->75.0->100.0 orders Step C ~ to Step F repeat.
H.
Finally the ‘SAVE’ button to save.
I.
Press the ‘EXIT’ button and get out.
The information displayed in ‘setting the position of UV’ screen is as follows:
Содержание HLM-7000P
Страница 1: ......
Страница 2: ...Auto Lensmeter HLM 7000P 1 Service Manual Auto Lensmeter HLM 7000P...
Страница 23: ...22 AUTO LENSMETER 3 Repair Standard 3 1 Removing Cover Assembly P I c t u r e...
Страница 25: ...24 AUTO LENSMETER 3 2 Frame Assemblely P I c t u r e...
Страница 27: ...26 AUTO LENSMETER 3 3 Back Bone Assembly P I c t u r e...
Страница 29: ...28 AUTO LENSMETER 3 3 1 PD Bar Sub Assembly P I c t u r e...
Страница 31: ...30 AUTO LENSMETER 3 3 2 Foot Frame Sub Assembly P I c t u r e...
Страница 33: ...32 AUTO LENSMETER 3 3 3 PEN Frame Sub Assembly P I c t u r e...
Страница 35: ...34 AUTO LENSMETER 3 3 4 PEN Slide Sub Assembly P I c t u r e...
Страница 37: ...36 AUTO LENSMETER 3 3 5 PD Pinion Sub Assembly P I c t u r e...
Страница 39: ...38 AUTO LENSMETER 3 3 6 LED Bone Sub Assembly P I c t u r e...
Страница 41: ...40 AUTO LENSMETER 3 4 LCD Assembly P I c t u r e...
Страница 43: ...42 AUTO LENSMETER 3 5 UV Bone Assembly P I c t u r e...
Страница 45: ...44 AUTO LENSMETER 3 6 Front Cover Assembly P I c t u r e...
Страница 48: ...Auto Lensmeter HLM 7000P 47...
Страница 56: ...Auto Lensmeter HLM 7000P 55 C Set the connected COM port and baud rate 115200 D Press Connect button...