PD Setup
With the PD Sensor, we can calibrate the PD value of our machine.
The procedures are as followings:
A.
Press the ‘PD’ (
) button.
B.
Push in the PD sensor is the leftmost. And press the ‘L Lim’ button.
C.
Push in the PD sensor is the rightmost. And press the ‘R Lim’ button and check if
‘PD POSITION[0] ~ PD POSITION[8]’ values are changed.
D.
Finally the ‘SAVE’ button to save.
The information displayed in ‘setting the position of PD’ screen is as follows:
Содержание HLM-7000P
Страница 1: ......
Страница 2: ...Auto Lensmeter HLM 7000P 1 Service Manual Auto Lensmeter HLM 7000P...
Страница 23: ...22 AUTO LENSMETER 3 Repair Standard 3 1 Removing Cover Assembly P I c t u r e...
Страница 25: ...24 AUTO LENSMETER 3 2 Frame Assemblely P I c t u r e...
Страница 27: ...26 AUTO LENSMETER 3 3 Back Bone Assembly P I c t u r e...
Страница 29: ...28 AUTO LENSMETER 3 3 1 PD Bar Sub Assembly P I c t u r e...
Страница 31: ...30 AUTO LENSMETER 3 3 2 Foot Frame Sub Assembly P I c t u r e...
Страница 33: ...32 AUTO LENSMETER 3 3 3 PEN Frame Sub Assembly P I c t u r e...
Страница 35: ...34 AUTO LENSMETER 3 3 4 PEN Slide Sub Assembly P I c t u r e...
Страница 37: ...36 AUTO LENSMETER 3 3 5 PD Pinion Sub Assembly P I c t u r e...
Страница 39: ...38 AUTO LENSMETER 3 3 6 LED Bone Sub Assembly P I c t u r e...
Страница 41: ...40 AUTO LENSMETER 3 4 LCD Assembly P I c t u r e...
Страница 43: ...42 AUTO LENSMETER 3 5 UV Bone Assembly P I c t u r e...
Страница 45: ...44 AUTO LENSMETER 3 6 Front Cover Assembly P I c t u r e...
Страница 48: ...Auto Lensmeter HLM 7000P 47...
Страница 56: ...Auto Lensmeter HLM 7000P 55 C Set the connected COM port and baud rate 115200 D Press Connect button...