2.2. 

  How to Set Origin 

 

The steps of setting origin are as follows:   

A. 

Press the ‘LENS’ (Key 1

)

 button.   

B. 

Press the ‘MEASURE’ (Key 2) button. 

C. 

Determine the lens to measure by using two ‘-INDEX’, ’+INDEX’ buttons. 

D. 

‘D-0.01’  button  and  ‘D+0.01’  button  by  the  given  value  in  the  Standard  Lens-set 

specification. 

E. 

To save and exit, press the ‘SAVE’ button. 

F. 

Press the ‘EXIT’ button again. And check if the diopter that has been set up in the 

basic display screen is shown.   

 

 

 

 

 

 

  ‘SAVE SCREEN INFORMATION’ is as follows: 

 

If  you  want  to  return  to  the  previous  one  without  saving  in  the  SAVE  screen,  press  the 

rightmost ‘PREV’ button. 

CEN: The x, y, coordinates of big center point.   
LEN: Average distance between of four points.   

CON: The x, y, coordinates of four big points.   
DIP: Saving order and the saved diopter.   

Buttons  for  Selecting 

Diopter

 

Buttons  for  Calibrate 

Diopter

 

SAVE button

 

Previous  screen 

button 

 

Содержание HLM-7000P

Страница 1: ......

Страница 2: ...Auto Lensmeter HLM 7000P 1 Service Manual Auto Lensmeter HLM 7000P...

Страница 3: ...ALIBRATION OF THE CYLINDRICAL AXIS 13 2 6 PD SETUP 15 2 7 UV SETUP 16 2 8 DIOPTER SETTING VARIABLES LIST 17 2 9 HOW TO CHANGE THE CCD CAMERA 18 2 10 HOW TO CHANGE THE LED ASSEMBLY 20 2 11 HOW TO CHANG...

Страница 4: ...BLOCK DIAGRAM 46 4 1 ELECTRICAL BLOCK DIAGRAM 46 4 2 INSPECTING SMPS 48 4 3 LCD TEST 49 4 4 LED TEST 49 4 5 UV MODULE TEST 50 4 6 PD TEST 51 4 7 THERMAL PRINTER TEST 52 5 HOW TO UPGRADE THE FIRMWARE S...

Страница 5: ...LENSMETER 1 Introduction 1 1 Components List Figure 1 Components Names I LCD Screen MEM Button Save Button Lens Table Lever Marking Lever Lens Holder Lever Printer Cover UV Cover Lens Cap Button Lens...

Страница 6: ...Auto Lensmeter HLM 7000P 5 Figure 2 Components Names II Figure 3 Components Names III Power Switch Power Connector RS 232 Connector...

Страница 7: ...N 1 Examining the Internal Components First 2 Inspecting The value S C A Prism Checking Abnormal State Cleaning the Pinhole Power off and on Checking Abnormal State Setup Again Exit Normal State Norma...

Страница 8: ...1 5 Optical System Optical system resolves the features of the lens such as diopter and prism It is composed of light emitting part and signal receiving part light emitting part composed of optic LED...

Страница 9: ...has transmitted the collimator lens passes the pinhole and makes image on CCD In case of convex lens lens with plus diopter parallel light that has transmitted the collimator lens converges through t...

Страница 10: ...DE mark will show in the upper area Calibration can be processed only in SETUP mode LENS PRISM CYLINDER PD UV To all setting you should convert the display step STEP into 0 01 In case of prism convert...

Страница 11: ...VE button F Press the EXIT button again And check if the diopter that has been set up in the basic display screen is shown SAVE SCREEN INFORMATION is as follows If you want to return to the previous o...

Страница 12: ...Green Color Cross Line will appear C Press the LENS Key 1 button D Press the MEASURE button E Determine the lens to measure by using two INDEX INDEX buttons F D 0 01 button and D 0 01 button by the g...

Страница 13: ...y the given value in the Standard Lens set specification E Rotate it to become the center position of 0 30 A0 degree F Press the MEASURE button and check if A0 A1 A2 A3 A4 A5 A6 A7 A8 A9 A10 A11 value...

Страница 14: ...nter 7 211 240 center 8 241 270 center 9 271 300 center 10 301 330 center 11 331 360 center 2 5 The calibration of the cylindrical axis With the 5D rectangular cylindrical lens we can calibrate the cy...

Страница 15: ...s set specification D Locate it to the center while cling it to the lens Axis 90 Prism_x y 0 E Press the MEASURE button and check if A90 value is changed F Finally the SAVE button to save CYLINDER 5Di...

Страница 16: ...ings A Press the PD button B Push in the PD sensor is the leftmost And press the L Lim button C Push in the PD sensor is the rightmost And press the R Lim button and check if PD POSITION 0 PD POSITION...

Страница 17: ...ecification E When UV transmissivity is 0 0 Prevent part that UV LED comes out When UV transmissivity is 25 0 50 0 75 0 Put a lens on the top of the UV LED comes out When UV transmissivity is 100 0 No...

Страница 18: ...erage distance and distance rates between the four points and 25D 1 19 99 L 168 52 119 09 119 23 average distance and distance rates between the four points and 20D 6 0 00 L 135 93 96 11 96 12 CX 642...

Страница 19: ...sembly C Detach the protective seal from the new CCD D Place the new CCD in its position Lift it up as much as you can But Be careful not to collide with the bone metal E Fasten the Base Screw I into...

Страница 20: ...MOS Camera Screws J If the sum of two A12 values in 6th row is greater than 0 1 loose all set screws for the Pinhole Housing And then rotate a little so that the sum of two A12 values is less than 0 1...

Страница 21: ...LED assembly The procedures for changing the upper frame Assembly are as followings A Remove the Side Cover Front Upper Cover and Upper Frame B Place the new Upper Frame and fasten it C Assemble the...

Страница 22: ...ole Housing and replace it the new one 6 Turn on the machine to the Setup Mode and then press LENS Key 1 button for a while to go into the Calibration Display 7 Press the first ALIGN button to see the...

Страница 23: ...22 AUTO LENSMETER 3 Repair Standard 3 1 Removing Cover Assembly P I c t u r e...

Страница 24: ...y Cables on Main Board And remove screw and plate on the LCD cable way Next removes 4 screws on LCD hinge and extract LCD Ass y Be careful that you never scratch on the surface of Cover while assembl...

Страница 25: ...24 AUTO LENSMETER 3 2 Frame Assemblely P I c t u r e...

Страница 26: ...ME CR RUBBER 1 3 FRAME CR RUBBER 2 4 FOOT 5 POWER SW BRACKET 6 B6_POWER SW SHEET 7 POWER SWITCH 8 LCD CABLE FIX BRACKET 9 MAIN BD BRACKET 10 B6_MB SPONGE GASKET 11 B6_MB EMI TAPE 12 B6_CAMERA SPONGE G...

Страница 27: ...26 AUTO LENSMETER 3 3 Back Bone Assembly P I c t u r e...

Страница 28: ...onents when assembling or disassembling the PD assembly Assembly is the reverse procedure of disassembly 2 B6_LED BONE ASSY 3 SHAFT FIX BRACKET 4 GUIDE SHAFT 4 5 B4_FOOT EXTENSION SPRING 6 FOOT MOVING...

Страница 29: ...28 AUTO LENSMETER 3 3 1 PD Bar Sub Assembly P I c t u r e...

Страница 30: ...ng for disassemble as needed Be careful that there is no scratch and assembly is the reverse procedure of disassembly 2 B4_PD SHAFT 3 PD RACK GEAR SHAFT 4 3 10 5 Z_WASHER_4X16 6 SEMS 3 10 7 O RING PD...

Страница 31: ...30 AUTO LENSMETER 3 3 2 Foot Frame Sub Assembly P I c t u r e...

Страница 32: ...nent Removal method 1 FOOT HOLDER Please see the drawing for disassemble as needed Assembly is the reverse procedure of disassembly 2 F T2 CONTROL POS PIN 3 POSITION FOOT SPRING 4 C7_POSITION FOOT RUB...

Страница 33: ...32 AUTO LENSMETER 3 3 3 PEN Frame Sub Assembly P I c t u r e...

Страница 34: ...g for disassemble as needed Assembly is the reverse procedure of disassembly 2 B4_PEN PRESS BRK 3 B4_PEN CONT SUPPT 4 2 6 6 5 SS 3 12 3 6 WRENCH BOLT 7 PEN GUIDE SHAFT 8 PEN GUIDE SPRING 9 E RING 10 2...

Страница 35: ...34 AUTO LENSMETER 3 3 4 PEN Slide Sub Assembly P I c t u r e...

Страница 36: ...o Component Removal method 1 B4_PEN SLIDE BASE Please see the drawing for disassemble as needed Be careful the balls Assembly is the reverse procedure of disassembly 2 B4_SLIDE ASSY 3 PCB SUPPORT BOLT...

Страница 37: ...36 AUTO LENSMETER 3 3 5 PD Pinion Sub Assembly P I c t u r e...

Страница 38: ...oval method 1 PD PINION BRK Please see the drawing for disassemble as needed Assembly is the reverse procedure of disassembly 2 PD PINION SHAFT 3 PD PINION NUT 4 PD PINION SHAFT HOLDER 5 GEAR WASHER 6...

Страница 39: ...38 AUTO LENSMETER 3 3 6 LED Bone Sub Assembly P I c t u r e...

Страница 40: ...t Removal method 1 B4_LED BONE 13 Turn part 2 for disassemble lens 3 Please see the drawing for disassemble as needed Assembly is the reverse procedure of disassembly 2 B4_LED LENS NUT 13 3 NEW_LM_LEN...

Страница 41: ...40 AUTO LENSMETER 3 4 LCD Assembly P I c t u r e...

Страница 42: ...separate 8 6 5 step by step Be careful that there is no scratch and assembly is the reverse procedure of disassembly 2 B6_MB EMI TAPE 3 B6_LCD SHIELD A 4 B6_LCD SHIELD B 5 B6_LCD WINDOW 6 LCD GRAPHIC...

Страница 43: ...42 AUTO LENSMETER 3 5 UV Bone Assembly P I c t u r e...

Страница 44: ...te 5 3 Remove 4 screws from 8 and separate 8 2 Be careful that there is no scratch and assembly is the reverse procedure of disassembly 2 UV TEFRON L 3 UV TEFRON U 4 HARNESS MB TO UV BLUE LED 5 PCB AS...

Страница 45: ...44 AUTO LENSMETER 3 6 Front Cover Assembly P I c t u r e...

Страница 46: ...ER TOP 5 Disconnect cables from 4 Remove 4 screws from 1 and separate 2 3 Be careful that there is no scratch and assembly is the reverse procedure of disassembly 2 COVER MEM BUTTON PLATE 3 COVER MEM...

Страница 47: ...4 1 Electrical block diagram CN6 CN2 CN3 CN1 LCD E05 CN11 CN4 CN3 BD1 H01 H03 H13 H10 H07 H08 H09 H06 BD4 BD2 CN1 CN1 E02 CN2 1 1 4 3 2 1 2 E N 1 4 3 L N L E01 E04 H02 H05 E03 CN22 CN21 CN14 H04 CN10...

Страница 48: ...Auto Lensmeter HLM 7000P 47...

Страница 49: ...Check the voltage of DC POWER Cable with DMM It should be AC100 220V REFERENCE ELECTRICAL CONSTRUCTION DIAGRAM Check1 Check BUZZER beeps after 2 seconds when you turn on the power switch MODULE TEST...

Страница 50: ...there is no character or figure Check2 There are horizontal lines on LCD and screen display is malfunctioned Check if connector of LCD is connected correctly to MAIN BOARD YES YES Change LCD Check1 P...

Страница 51: ...eck if light radiated from OPTIC LED comes out from UV measurement part NO Change UV LED PCB ASS Y Check1 Check if UV SENSER in UV measurement part is hidden by obstacles such as FILTER COVER NO Check...

Страница 52: ...o Lensmeter HLM 7000P 51 4 6 PD TEST PD value is not displayed Check1 Check right left switching when you move LEVER right and left compared to the center of PD NO Change MAIN BOARD Change PD MODULE Y...

Страница 53: ...roperly Check1 Open PRINTER COVER and check if HEADER LEVER towards upside Check1 Open PRINTER COVER Check2 Check PRINTER PAPER tray Check3 Fill up PRINTER PAPER if PRINTER PAPER is empty or near empt...

Страница 54: ...comes out If there is PC with Windows operating system and serial cable connected to the Lensmeter with simple process about 3 minutes he she can upgrade the performance of his her Lensmeter without s...

Страница 55: ...54 AUTO LENSMETER 5 2 How to upgrade the firmware software using DNW application in HLM machines A Execute the DNW application B Press Options button to set the COM port...

Страница 56: ...Auto Lensmeter HLM 7000P 55 C Set the connected COM port and baud rate 115200 D Press Connect button...

Страница 57: ...56 AUTO LENSMETER E Power on the Lensmeter with pressing 5 key Then you can see the below messages F Press Transmit button and select firmware file...

Страница 58: ...Auto Lensmeter HLM 7000P 57 G Wait until download is the end H Reboot the Lensmeter after restart system message...

Страница 59: ...m 0 10 0 01 0 125 0 25 0 20 0 01 0 06 0 12 0 25 Measurable lens diameter 15 to 115mm 15 to 120mm Method 4 point Hartmann Sensor with 81 Multiple Measuring Points UV 0 100 0 100 Design Size 190 W X237...