![Huvitz HLM-1 Скачать руководство пользователя страница 9](http://html1.mh-extra.com/html/huvitz/hlm-1/hlm-1_service-manual_2190264009.webp)
8
AUTO
-
LE
NSMETE
R
Signal Receiving Part
-
Hartmann plate : divide the signal into 81 points. Prevent the other light except the
signal
-
2-dimentional CMOS sensor
: converts signals into digital value.
1.6.
Measurement Principle
Basic Principle
When there is no lens (0D state), parallel light that has transmitted the collimator lens
passes the pinhole and makes image on CCD. In case of convex lens (lens with plus
diopter), parallel light that has transmitted the collimator lens converges through the
measuring lens, passes the pinhole and then makes image on CCD. On the other hand, in
case of concave lens (lens with minus diopter), parallel light that has transmitted the
collimator lens diverges through the measuring lens and then passes the pinholes and at
last makes image on CCD.
2.
Initial State
Concave Lens
Convex Lens
Содержание HLM-1
Страница 1: ......
Страница 2: ...Auto Lensmeter HLM 1 1 Service Manual Auto Lensmeter HLM 1...
Страница 21: ...20 AUTO LENSMETER 3 Repair Standard 3 1 Removing Cover Assembly P I c t u r e...
Страница 23: ...22 AUTO LENSMETER 3 2 Frame Assemblely P I c t u r e...
Страница 25: ...24 AUTO LENSMETER 3 3 Back Bone Assembly P I c t u r e...
Страница 27: ...26 AUTO LENSMETER 3 3 1 PD Bar Sub Assembly P I c t u r e...
Страница 29: ...28 AUTO LENSMETER 3 3 2 Foot Frame Sub Assembly P I c t u r e...
Страница 31: ...30 AUTO LENSMETER 3 3 3 PEN Frame Sub Assembly P I c t u r e...
Страница 33: ...32 AUTO LENSMETER 3 3 4 PEN Slide Sub Assembly P I c t u r e...
Страница 35: ...34 AUTO LENSMETER 3 3 5 PD Pinion Sub Assembly P I c t u r e...
Страница 37: ...36 AUTO LENSMETER 3 3 6 LED Bone Sub Assembly P I c t u r e...
Страница 39: ...38 AUTO LENSMETER 3 4 LCD Assembly P I c t u r e...
Страница 41: ...40 AUTO LENSMETER 3 5 Under Cover Assembly P I c t u r e...
Страница 44: ...Auto Lensmeter HLM 1 43...
Страница 49: ...48 AUTO LENSMETER C Set the connected COM port and baud rate 115200 D Press Connect button...
Страница 51: ...50 AUTO LENSMETER G Wait until download is the end H Reboot the Lensmeter after restart system message...