Excelon EZ Operator’s Manual
Size Calibration
Procedures for size calibration
①
Replace Lens adaptor
holder with Standard JIG.
②
Press the [F1] button to
start size calibration.
③
After calibration, press the
[F1] button to save or [F2]
button to discard.
Feeler Calibration
Same as above size calibration procedure except step
②
.
Press the [F2] button to start feeler calibration.