7. Connect the garment(s) to the two 12-way connectors on the pump, as
follows:
•
If two garments are used (either two arm garments, two leg
garments, or one arm/leg garment and one garment insert), one will
be connected to the upper connector and the other connected to the
lower connector. The two connections are interchangeable - both
garments will inflate at the same time and to the same pressure.
•
If only one garment is used, it must be connected to the upper
connector and the rubber cover must be folded over and securely
fitted to the lower connector.
Note:
The connectors are push-fit and polarised to prevent incorrect
orientation.
8. If the patient is setting up or operating the system on his/her own, make sure
that:
•
The pump is set up
before
garments are fitted.
•
The garments are connected to the pump
before they are fitted onto the limbs.
9. Insert the mains power plug into a suitable mains power socket.
10. Switch on the mains power supply to the pump.
11. The pump will perform a short self-test and then go into Standby. Self-test
may take between 15 to 20 seconds.
12. The Run and Sleep mode indicators are extinguished.
13. The pressure display on the pump control panel will show the previously
selected pressure setting.
14. Check the therapy setting display reads 35.