MiniScan EZ Maintenance and Testing
The MiniScan EZ does require some maintenance. This chapter outlines the parts of the MiniScan EZ
you must maintain in order for the instrument to function properly and tests that you may run to
assess its performance.
Note: The MiniScan EZ contains hazardous voltages and no user-replaceable parts. It should be
disassembled only by HunterLab personnel.
Running the Repeatability Test
You may test the repeatability of your instrument as follows:
1.
Turn the MiniScan EZ on and allow it to warm it up for 2 hours. Meanwhile, clean the white tile
as described on the next page and allow the tile to return to room temperature.
2.
Follow the instructions given in the
Sensor Menu > Diagnostics
section to run the repeatability
test that is built into EasyMatch QC.
3.
Standardize the instrument. Place the white tile flush at the port and read 20 times by pressing
Yes
.
Figure 16. In Progress Repeatability Test
4.
When the test is completed, a report is generated as shown in the next Figure.