ColorFlex EZ Citrus Citrus
User’s Manual Version 2.0
29
Figure 44. Standard Values
For a
PHYSICAL
standard, place the standard at the sample port and press the
GO
button (Read)
to read the standard and place its measurement values (in the selected color scale) in the
product setup. Then press the
LEFT ARROW
button (Done) to return to the product setup
configuration screen.
For a
NUMERIC
standard, change the highlighted digit of the selected color scale value using
the
UP
and
DOWN ARROW
buttons. When the desired digit is shown, press
NEXT
to move to
the next digit. When all digits are set as desired, press
DONE
to return to the product setup
configuration screen.
Figure 45. Enter a Numeric Standard
For a
HITCH
standard, place the hitch standard (with known values from a reference
instrument) at the sample port and press the
GO
button (Read) to read the standard and place
its measurement values (in the selected color scale) in the product setup.
Figure 46. Hitch Standard
Then, change the highlighted digit of the selected color scale value to match the corresponding
digit of the known values from the reference instrument using the
UP
and
DOWN ARROW
buttons. When the desired digit is shown, press
NEXT
to move to the next digit. When all digits
match the corresponding digit from the reference instrument’s reading, press
DONE
to return
to the product setup configuration screen.
The offsets between the measured and entered values for the hitch standard are saved with
the standard and applied to all samples read in that setup.
Содержание ColorFlex EZ Citrus
Страница 6: ...ColorFlex EZ Citrus Citrus User s Manual Version 2 0 6...
Страница 18: ...ColorFlex EZ Citrus Citrus User s Manual Version 2 0 18...
Страница 44: ...ColorFlex EZ Citrus Citrus User s Manual Version 2 0 44...
Страница 54: ...ColorFlex EZ Citrus Citrus User s Manual Version 2 0 54 Regulatory Notice...
Страница 60: ...ColorFlex EZ Citrus Citrus User s Manual Version 2 0 60...