13 of 35
01115009A _9
Paracube
®
Micro – Instruction Manual
4.2
Electrical Arrangement
Power Supply (to be provided by the OEM)
The sensor requires an external power supply as specified in Section 3.3.
Electrical Connection
Connection to the sensor is made via a 4 way connector mounted onto the sensor’s
PCB. There are two connector types offered as standard, a Molex KK type friction lock
(Fig. 5) or a low profile SMT connector (Fig. 6). See Table 1 for pin out details. Full
details of the sensor variants are detailed in Section 6.1.
All electrical connections to the sensor must be made using the correct style of ‘Molex’
connector, website www.molex.com
PCB Mounted KK Friction Lock Connector - Molex Part No 0022272041, Fig. 5
2.54mm (.100") Pitch KK® Wire-to-Board Header, Vertical, with Friction Lock.
Mating crimp housing required by end-user Molex part number
0022012045
used in
conjunction with crimp terminal part number
0008500032.
Fig 5
CAUTION – No. 4
The four screws securing the outer most PCB are factory fitted and
should not be removed or used for mounting purposes
WARNING USE - No. 1
F
AILURE TO FOLLOW THE RECOMMENDED PROCEDURE FOR FIXING OF THE
S
ENSOR MAY RESULT IN LEAKS EXPOSING PERSONNEL TO THE SAMPLE
GASES
.
Содержание Paracube Micro
Страница 2: ...2 of 44 01120001A _9 Paracube Modus Instruction Manual This page is intentionally blank...
Страница 31: ...28 of 35 01115009A _9 Paracube Micro Instruction Manual Appendix 7 3 Procedure for Fixing of the Adaptor Plate...
Страница 36: ...33 of 35 01115009A _9 Paracube Micro Instruction Manual Appendix 7 6 RoHS II Directive 2011 65 EU Declaration...
Страница 38: ...35 of 35 01115009A _9 Paracube Micro Instruction Manual...