45
5. When the Deairing Tank is 3/4 full, turn the Deairing Tank
FILL/DRAIN valve to the “OFF” position.
6. Leave the VACUUM/PRESSURE valve on VACUUM for 20-30
minutes to remove air out of water in Deairing Tank. The Vacuum
should be in the -28” HG. Then…
7. After 30 minutes, turn the VACUUM/PRESSURE valve to the OFF
position. The Valves should now be in the position shown in the
photo below
(3)
.
8. You can now turn off your Vacuum Pump
9. You can now turn the Lower Deairing Tank valves from Off to
Pressure
(4)
.
You should now have an adequate supply of Deaired Water in
the Deairing Tank to initiate a triaxial test.