34
Note: An input cannot be calibrated during testing.
Units (2)
In this field, a calibrated instrument will display the units that were
chosen for use at the time of calibration. This field can also be used
to automatically toggle conversion of units between lb.-in. and SI
units if the need arises. To view choices for types of Units, click on
the Units field and the Unit choices will be displayed, see image
below.
Value (3)
This field displays the current calibration value. This value should
already be set with 3 decimal point accuracy . If the instrument is not
calibrated, the unit will read “N/A”. Clicking on this field will bring up
a window where you can enter the required decimal points required
for the value.
Each channel has a “Limits On” check box (1). Use the Limits On to
keep the machine from exceeding the sensor limits of the instrumen-
tation. By selecting this option, before the test can exceed the limits
of the sensors, all tests will stop running and the motor will stop to
avoid damaging connected instrumentation.