product manual 07.22

Automated Pressure Controller

with Load Frame

HM-5820.3F

Содержание HM-5820-150.3F

Страница 1: ...product manual 07 22 Automated Pressure Controller with Load Frame HM 5820 3F ...

Страница 2: ...st Templates 20 Save to USB Check Box 20 Initial Set Up Network 20 Network Settings Screen 21 DHCP 21 IPInformation 22 Local Status 22 InternetStatus 22 Initial Set Up Information 22 SystemInformation 23 FirmwareVersion 23 IPInformation 23 LocalStatus 23 InternetStatus 23 Memory 23 FactoryScreen 23 ExportLogFile 23 Initial Set Up Contact 24 ContactInformation 24 Initial Set Up Update 25 Updatefrom...

Страница 3: ...lates 20 Save to USB Check Box 20 Initial Set Up Network 20 Network Settings Screen 21 DHCP 21 IPInformation 22 Local Status 22 InternetStatus 22 Initial Set Up Information 22 SystemInformation 23 FirmwareVersion 23 IPInformation 23 LocalStatus 23 InternetStatus 23 Memory 23 FactoryScreen 23 ExportLogFile 23 Initial Set Up Contact 24 ContactInformation 24 Initial Set Up Update 25 UpdatefromUSB 25 ...

Страница 4: ...th most power configurations throughout the world The unit is supplied with an IEC electrical cord with a standard 110V plug The HM 5820 arrives ready for operation Attach the sup plied IEC electrical cord to the machine and plug into a standard wall receptacle for use in the United States For locations other than the U S replace the supplied electrical cord with an IEC cord that has the correct p...

Страница 5: ... used for powering a wireless access appliance for those who want to use a wireless LAN setup Instrument Inputs 3 The panel features four 4 inputs for connecting instrumentation to the machine Each input represents a separate channel If you have purchased instrumentation from Humboldt with your HM 5820 it has been assigned to a specific channel and calibrated In strumentation will have a numbered ...

Страница 6: ...the machine up for use Please note that if you are using the HM 5820 in conjunction with Humboldt s NEXT testing software the calibrations will need to be imported into the software before use Below are photos of an instrumentation input and the instrumentation plug Install the plugs into the inputs by lining up the guide at the bottom of the plug with the slot at the bottom of the input Instrumen...

Страница 7: ...ur HM 5820 is first turned on the screen below will appear From this screen to confirm your machine instrumentation has been calibrated navigate to the Calibration section by clicking the Home icon in the top left corner of the screen When you click on this button you will see a drop down menu appear DO NOT RECALIBRATE Initial Set Up Home Button The Home button is the first step in navigating the ...

Страница 8: ...gate to the bottom of this drop down menu and click on System Settings 1 You will see the following screen Click on the Calibration tab in the top left corner 2 You will see the following screen 2 1 Initial Set Up Calibration ...

Страница 9: ...information A green box at the left of a channel indicates that the channel has instrumentation assigned to it and that it is calibrated and ready for use 1 1 2 3 The HM 5820 uses a total of eight inputs for testing On the main calibration screen this is divided into two screens for ease of viewing and input setting adjust ment To switch between the screen showing channels 1 4 and the screen showi...

Страница 10: ... before the test can exceed the limits of the sensors all tests will stop running and the motor will stop to avoid damaging connected instrumentation Units 2 In Units field a calibrated instrument will display the units of measure that were chosen for use at the time of calibration The units of measure can be toggled between English and SI units as desired Value 3 This field displays the current c...

Страница 11: ...ayed are as follows Input 1 Cell Pressure 0 200 psi Input 2 Back Pressure 0 200 psi Input 3 Pore Pressure 0 200 psi Input 4 Load 0 2 000 lbf The second channel series channels 5 8 are available when the toggle button is pressed The second series of channels are set as follows Input 5 Displacement 0 2 000 in Input 6 Cell Volume 0 250 cm Input 7 Base Volume 0 250 cm Input 8 Top Volume 0 250 cm ...

Страница 12: ...ate the vol ume the following steps should be followed 1 Ensure that the piston in the volume chamber to be cali brated has been moved down to the maximum extent 2 Select the HOME button then select CALIBRATIONS 3 Select the volume sensor to be calibrated Cell Base or Top by clicking the corresponding CALIBRATE button This will open the SENSOR DETAILS window 4 When prompted enter the calibration e...

Страница 13: ... 13 7 A pop up window will appear asking if you wish to reset the encoder counts Select YES 8 Select a name for the instrument being calibrated or you can leave it blank and the machine will use the factory default name ...

Страница 14: ...llows you to recover your calibration data quickly On the screen below press the Export Calibration button 1 to select calibrations to export via USB This thumb drive will contain the calibration data for the extra instrumentation To use this calibration information it is rec ommended to first export all your HM 5820 s calibration data to a thumb drive and then download the additional calibra tion...

Страница 15: ...k view either a 24 hour or 12 hour clock as well as the option to show seconds or not Clock 4 The current time is displayed is located on every controller screen Clicking on the time from any screen you will be taken to the Date Time Settings Screen 4 2 3 1 Click on the Date Time tab 1 You will see the following screen 1 Initial Set up Date and Time ...

Страница 16: ...ck the yellow box to change the number of min utes before the display goes dim After the time has elapsed touch the display to deactivate dimming Setting the value to 0 will turn off the specific function For example setting TURN DISPLAY OF to 0 will result in the screen remaining on indefinitely To set up Display settings return to the System Settings screen and click on the Display Panel 1 1 Ini...

Страница 17: ...e Logger ID 1 Each machine that is connected to your network requires a unique Logger ID These numbers can be assigned any Turn Display Off 3 The display will automatically turn off to save power Click the yellow box to change the number of minutes before the display powers off This screen is accessed by clicking on the Preferences button 1 1 Initial Set Up Preferences ...

Страница 18: ...System Units Tab This screen is accessed by clicking on the System Units Tab under System Preferences Ambient Temperature 1 This field displays the current ambient temperature and allows you to select desired units Fahrenheit or Celsius for temperature recording Motor Speed Unit 2 Click on the desired units you want to use for motor control in min or mm min Stress Control Unit 3 Click on the desir...

Страница 19: ...umber can be between 10 50 Specimen Height 1 This field allows you to set the specimen height that will be used for all testing calculations Specimen Diameter 2 This field allows you to set the specimen diameter that will be used for all testing calculations Preferences Storage Tab This screen is accessed by clicking on the Storage Tab un der System Preferences 1 2 ...

Страница 20: ...not displayed in the Previous Tests window These tests are not deleted they are recycled for later use Select this button to view ex port delete recycled tests Test Templates 3 Select this button to delete test templates that have been created but are no longer desired Save to USB Check Box 4 Checking this Box saves the results of a test to USB thumb drive inserted into the front USB slot on the m...

Страница 21: ...ings screen it pro vides information on your IP information and network status DHCP 1 Check this box to enable disable the Dynamic Host Config uration Protocol DHCP If enabled your machine will pick up IP information from your router If disabled you will need to manually enter the network information for a static IP please consult your network administrator for this ...

Страница 22: ...achine Local Status 3 This indicates the status of the local network connection Connected or Disconnected Internet Status 4 This indicates the status of your Internet connection Connected or Disconnected Note If you are experiencing issues with any connec tions please contact your IT department for assistance Clicking on this panel provides a view of the current status of the machine 1 Initial Set...

Страница 23: ... local network connection Connected or Disconnected Internet Status 4 This indicates the status of your Internet connection Connected or Disconnected Memory 5 This indicates the current status of how much available memory is being used by the machine Factory Screen 6 This is for Humboldt use only Export Log File 7 This button exports a log file from the machine to a USB thumb drive Be sure to inse...

Страница 24: ...upport and Service For quickest response go to this link on our website https www humboldtmfg com support and fill in the support form This will provide us with the necessary information to assist you and you will be added to the next position in the support cue You can also email Humboldt Support at support humboldtmfg com or Humboldt Service at service humboldtmfg com Please include contact info...

Страница 25: ...ps www humboldtmfg com sup port software php Once on this page click on the Elite Series Firmware tab You will see a list of Humboldt Elite Series machines Click 1 Clicking on this panel provides information on checking for Updates performing updates and an update history for the machine 1 1 Clicking on the Update Panel brings up the following screen Initial Set Up Update ...

Страница 26: ...e update to fail preventing the machine from functioning properly Check for Update 2 It is recommended that you update your HM 5820 firmware periodically Checking for firmware updates every other month should be sufficient to keep your system up to date on the HM 5820 Current Version link and the firmware update will begin to download to your computer Once the download is complete Load the file on...

Страница 27: ...Equipment Setup 27 ...

Страница 28: ...her than the U S replace the supplied electrical cord with an IEC cord that has the correct plug for your application The supplied cord can also be used by cutting the standard plug from the cord and attaching the correct plug Power Switch The Power Switch is located on the left side of the back of the machine above the electrical cord inlet The Fuse Compartment is located between the electrical c...

Страница 29: ...pecific channel and calibrated Instrumentation will have a numbered tag attached to it and should be plugged into the corresponding input on the machine Below are photos of an instrumentation input and the instrumen tation plug Install the plugs into the inputs by lining up the guide at the bottom of the plug with the slot at the bottom of the input Below are photos of an instrumentation input and...

Страница 30: ...ill have to import the correct calibration infor mation to that input for correct operation Calibration information for additional instrumentation can be found on a USB thumb drive that was shipped with your HM 5820 For information on calibration see Page 9 of this manual Instrumentation Setup On the next page are examples of typical instrumentation setups for different tests the HM 5820 is capabl...

Страница 31: ... with your HM 5820 has been cali brated and assigned to specific channels for use These channels are marked on the corresponding instrumentation and the correct calibration information will appear in the channel calibration pa rameters once you have installed the instrumentation to the correct inputs DO NOT RECALIBRATE How to Perform a Calibration All instrumentation included with this unit has be...

Страница 32: ...ibration section by clicking the Menu icon in the top left corner of the screen 1 When you click on this button you will see a drop down menu appear see below You will see the following screen 2 1 Navigate to the bottom of this drop down menu and click on System Settings 2 You will see the following screen ...

Страница 33: ...itor and cal ibrate instrumentation and assign them to specific channels of the HM 5820 The Calibration Input Screen provides a summa ry of the calibration status of each channel At this time verify the calibration information A green box at the left of a channel indicates that the channel has instrumentation assigned to it and that it is calibrated and ready for use 1 1 ...

Страница 34: ...Value 3 This field displays the current calibration value This value should already be set with 3 decimal point accuracy If the instrument is not calibrated the unit will read N A Clicking on this field will bring up a window where you can enter the required decimal points required for the value Each channel has a Limits On check box 1 Use the Limits On to keep the machine from exceeding the senso...

Страница 35: ...tion that is already being used for that input To do this Press the Export Calibration Button 4 see next page to select calibrations to export via USB It is a good practice to export all your calibrations to a thumb drive In case of a problem this practice al lows you to recover your calibration data quickly 3 4 5 ...

Страница 36: ... calibrations you wish to recalibrate When you press the trash can icon this screen will appear On this screen select an Input calibration to delete one at a time and then press the Select button 1 The calibration will be deleted You can do this for all Inputs you wish to calibrate Once all Input calibrations have been cleared your Calibration window should look like the one below with no Inputs c...

Страница 37: ...accurate The cal ibration process involves plugging the instrumentation into the HM 5820 while placing the instrumentation into the certified calibration device which provides a specific set load or displacement Once this has been done click the Calibrate button next to the appropriate Input 1 A pop up window will appear requiring you to enter the password which is 22234 see below 1 ...

Страница 38: ...the maximum capacity of the sensor and choose either cm3 or in3 For Counts per Unit 3 refer to the calibration sheet which came with the instrument you are using and enter it here and then choose mV V The next page shows typical calibration sheets with the FS Output information highlighted in orange NOTE For volume calibrations Counts Per Unit will always be 9295 680 1 2 3 4 38 ...

Страница 39: ...mum of 10 points Once this is complete click on the Right Arrow 4 in the bottom right hand corner of the screen to save these settings You will be taken to Tab 2 Limits 1 2 3 If you choose Per FS Output not a recommended calibration meth od it will base your calibration on the maximum capacity of the sensor vs the minimum capacity of the sensor If you choose this method click on the Right Arrow in...

Страница 40: ...on method Instead the calibration is determined by using a stepper motor to determine the volume of water currently in the pressure cylinder On Tab 3 you will be able to set the number of points you want to use for your calibration You can choose 1 point 5 point 10 point or Custom which allows you to use any number of points up to a total of 10 In the example on the previous screen shot above a 5 ...

Страница 41: ...essure rating of 200 psi you would set the 0 point at 0 and the 5 point at 200 Points 1 through 4 would typically be set at 1 40 psi 2 80 psi 3 120 psi 4 160 psi and 5 200 psi Or divide the maximum load number by the number of points in this case 5 which works out to 40 point increments between points See photo below Each point would be chosen by clicking on the corresponding calibra tion point ro...

Страница 42: ... disk icon in the lower right hand corner 1 This will save it Use this method for calibrating any additional Inputs necessary 1 On Tab 4 you will be asked to name the calibrated Input If a user de fined name is not chosen the machine will save the calibration using a default name based on the sensor type and channel input number ...

Страница 43: ...Test Setup 43 ...

Страница 44: ...begin set all knobs into the Off position see photo below Then 1 In the Deairing Tank section Turn the top FILL DRAIN valve to the FILL position 1 2 Then Turn the bottom VACUUM PRESSURE valve to the VACUUM position 2 The valves should now be in the position shown in the photo below 1 2 ...

Страница 45: ...Tank The Vacuum should be in the 28 HG Then 7 After 30 minutes turn the VACUUM PRESSURE valve to the OFF position The Valves should now be in the position shown in the photo below 3 8 You can now turn off your Vacuum Pump 9 You can now turn the Lower Deairing Tank valves from Off to Pressure 4 You should now have an adequate supply of Deaired Water in the Deairing Tank to initiate a triaxial test ...

Страница 46: ...ill need to be done three times once for each cylinder Cell Base and Top With the HM 5820 on you should see this screen 1 Press the FILL tab on the right side of the touch screen 1 2 Press the fill button which corresponds to the Cell Pressure Chamber on the HM 5810 5820 2 1 2 ...

Страница 47: ... way to the top of the cylinder close the bleed valve 3 above and press the down arrow on the touch screen and the piston will move to the bottom of the cylinder and read 0 0000 This will allow the piston to move down and fill the cylinder with water 3 4 3 Open the Cylinder Bleed valve 3 NOTE Make sure the Line Bleed valve is closed 4 5 ...

Страница 48: ...ter is exiting close the Bleed Valve 6 Open the Line Bleed valve to allow water to exit the lines and fittings near the pressure cylinder When all air has been removed through the Line Bleed valve close the bleed valve Press the Fill button on the touch screen to close the solenoid valve NOTE This process is typical for filling all three pressure cylinders 8 Repeat this process for the Base and To...

Страница 49: ... Touch Screen 3 Enter 50 psi for Cell Pressure in the yellow PSI Field 2 and then Press the corresponding green Start Button 3 see photo be low 10 2 Deairing the Lines from the HM 5810 5820 to the Triaxial Cell 1 Connect a length of tubing to the CELL PORT on the front of the HM 5820 2 see photo at top of page 2 Press the FILL BUTTON 10 on the touch screen for the CELL CHAMBER 3 Press STOP 2 on th...

Страница 50: ...50 3 Repeat this process for the Base and Top Chambers 4 Leave all three chambers set under 50 psi for 24 hours to dis solve any remaining air in the HM 5810 5820 plumbing system 3 1 2 ...

Страница 51: ... on the left front of the HM 5820 nd the Cell port on the Triaxial Cell NOTE There is insufficient water contained within the pressurec cylinders to fill the Triaxial cell Do not try to fill the interior lines or the Triaxial Cell using the water contained within the pressure cylinders 2 Ensure that the PRESSURE RELIEF VALVE is open on top of Triaxial Cell NOTE On Humboldt Triaxial Cells this is a...

Страница 52: ...closed see CELL DWG 2 above Remove the Vent Adapter to HM 5820 Cell Port Close Cell Valve CELL DWG 2 5 Once the Triaxial Cell is filled completely turn the CELL VALVE on the Triaxial Cell to the OFF position See CELL DWG 2 below Press the FILL tab on the touch screen to close the water supply 6 Then Remove the Vent Adapter at the top of the Triaxial Cell See CELL DWG 2 below 1 ...

Страница 53: ...port to the right of the CELL VALVE on the Triaxial Cell see CELL DWG 3 below 3 In addition connect a line between the TOP VALVE and the BASE VALVE on the left side of the Triaxial Cell see CELL DWG 3 below BASE PORT HM 5820 Photo TOP PORT CELL PORT to HM 5820 Base Port to HM 5820 Top Port to HM 5820 Cell Port ...

Страница 54: ...e air tank through the HM 5820 and through the internal lines on the Triaxial Cell Note For this operation to work properly the De airing tank must be positioned at least 2 3 feet above the HM 5820 and the Triaxial Cell Monitor the base pressure on the HM 5820 and ensure that the Base pressure does not exceed the Cell pressure during this operation 8 When all the air has been bled from the interna...

Страница 55: ...esponding green START BUTTON 4 below CAUTION ENSURE THAT THE CELL PRESSURE SETTING IS ALWAYS GREATER OR EQUAL TO THE BASE AND TOP PRESSURES ALSO ENSURE PRESSURE IS APPLIED TO CELL BEFORE BASE AND TOP FAILURE TO DO THIS MAY RESULT IN OVERFILLING THE SAMPLE WITH WATER AND RESULT IN TEST FAILURE 4 ...

Страница 56: ...an be set from 1 cc sec 60 cc min down to less than 0 000002 cc sec 0 00012 cc min There are three data input channels one for each pressure transducer The HM 5820 provides an accurate and simplified permeability setup which eliminates the need for separate distribution panels and simplifying tubing and control cable setup By using the integral staging platform the HM 5810 provides an extremely co...

Страница 57: ...lite Series load frames the HM 5820 is built around Hum boldt s integral 4 channel data logger with touch screen control which allows the load frame to be used as a standalone device capable of full test con trol and data logging It can also be controlled by a networked computer at any location with access to the network Controller Specifications Display Resistive Touch 7 178mm VGA 480 x 800 Real ...

Страница 58: ...y or subsequently sold Fitness for Application The manufacturer makes no recommendations or claims regarding fitness for appli cations other than the specific tests as defined in this User Guide Unpacking Initial inspection should include checking for physical damage during shipping and obvious external damage to the product Package contents are defined by your packing list Each Loader is configur...

Страница 59: ...ption of HUMBOLDT Authorized Service Facilities you may not copy alter decompile or reverse assemble the software in any fashion except as instruct ed in this manual US copyright laws trademark laws and trade secrets protect the materials Any person s and or organizations that attempt or accomplish the above violation or knowingly aid or abet the violation by supplying equipment or technol ogy wil...

Страница 60: ...Humboldt Mfg Co 875 Tollgate Road Elgin Illinois 60123 U S A U S A Toll Free 1 800 544 7220 Voice 1 708 456 6300 Fax 1 708 456 0137 Email hmc humboldtmfg com ...