26
NOTE: If you apply all the surcharge weight in the mold at the beginning
of the test, just Press
to continue. The HM-3000 will start the test.
The display will show the current reading of the load and displacement.
The HM-3000 will collect a load reading every 0.025 inches. Once the
displacement reaches 0.5 inches, the machine will stop collecting data
and reverse the platen back to the lower limit at 2.0 in/min (50.8 mm/min).
At any time you can press
to end the test. This will stop the data
collection and reverse the platen to its down position. You should see this
screen:
Pressing
(PRINT DATA) will print the data to a printer or save a file
in ASCII mode to your computer using Humboldt Material Testing Software
(HMTS) or Windows Hyper terminal. Once the data has been printed or
transferred, you can begin another test after you have readied a test mold on
the platen by pressing
(RUN CBR/LBR TEST).
If you don’t have a printer or a computer connected to the HM-3000, you
can press
(SCROLL DATA) to scroll the data. You should see this
screen:
Using the
(NEXT) key to scroll through the readings and the
(LAST) key to view the last reading taken.
To return to any prior screen you can always press
Marshall
Center the breaking head on the platen. Press the
button to raise the
head to the start position, close to the load cell. There should be about
a 1/8" (3.000mm) gap between the breaking head and the load cell.
Press
You should see this screen:
* PRINT DATA
* SCROLL DATA
* RUN CBR/LBR TEST
LOAD
“READING”
PENETRATION
“READING”
* NEXT
* LAST
READING #
Содержание Digital MasterLoader HM-3000.3F
Страница 1: ...product manual 07 08 HM 3000 3F Humboldt Digital MasterLoader ...