12.0 RADIATED EMISSIONS
The use of accessories and cables, other than those specified or sold by the manufacturer of the equipment as
replacement parts for internal components, may result in increased emissions or decreased immunity of the equipment.
The T7 Point-of-Care Technology Cart may cause interference when used adjacent to or stacked with electrical
equipment outside the system and could affect normal operation of said equipment. Do not stack additional electrical
equipment on top of the work surface enclosure.
The T7 Point-of-Care Technology Cart is intended for use by healthcare professionals only. It may cause radio
interference or may disrupt the operation of nearby equipment. It may be necessary to take mitigation measures, such
as re-orienting or relocating the T7 Cart or shielding the location.
12.1 RECOMMENDED SEPARATED DISTANCES
The T7 is intended for use in the electromagnetic environment specified below in which radiated disturbances are
controlled. The T7 user can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF Communications Equipment and the T7 as recommended below, according to the maximum
output power of the communications equipment.
WARNING: Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30cm (12 inches) to any part of the T7, including cables specified by
Humanscale. Otherwise, degradation of the performance of this equipment could result.
For transmitters rated at a maximum output power not listed above, the recommended separation distance (D) in meters
(M) can be determined using the equation applicable to the frequency of the transmitter, where (P) is the maximum
output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Max Output Power
(W)
Separation
(M)
Separation
(M)
Separation
(M)
150 kHz to 80 MHz
D = 1.2 (SartP)
81
to 800 MHz
D
= 1.2 (Sart P)
801 MHz
to 2.5 GHz
D
= 1.2 (Sart P)
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
Equipment and Systems that are NOT Life-Supporting
Note 1: At 80 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
56
Содержание TouchPoint T7
Страница 1: ...TouchPoint T7 User Manual ...
Страница 3: ...INTRODUCTION 2 Return to Table of Contents ...
Страница 28: ...Calculator Enabling Calculator Enabled Calculator Disabled Swipe 27 Return to Table of Contents ...
Страница 34: ...GETTING STARTED 33 Return to Table of Contents ...
Страница 38: ...OPERATION 37 Return to Table of Contents ...
Страница 47: ...TECHNICAL INFORMATION 46 Return to Table of Contents ...
Страница 56: ...11 1 MONITOR ADJUSTMENT RANGE Rotation Range Tilt Range 15º 90º 90º 55 Return to Table of Contents ...