ELISYS QUATTRO
- User Manual HUMAN
System Basics
2.2.2
Patient Sample and Reagents Unit with Bar
Code Reader
CAUTION: Laser radiation - do not stare into beam!
The bar code scanner is located on the right side at the front of the patient sample
and reagent unit.
The patient sample and reagent unit, accessible via a flap, includes 12 tracks for
sample and reagent racks.
The following racks are supplied:
Each rack includes a contact pin; on racks occupying one track, this pin is located at
the top centre, and on the broader racks at the top right.
The software specifies which track is to be used for the respective rack. This is indi-
cated by a red LED. A reagent rack occupying 3 tracks must be inserted such that
the contact tappet is in contact with the lit up LED.
Each rack has to be inserted up to the limit stop. The respective LED on the rear pan-
el goes off
(Loading/Unloading see chapter 5 on page 5-1)
.
Reloading of patient and reagent racks is possible when the instrument is in the in-
cubation mode.
Keep the flap of this unit closed during a run. It may be opened only for re-
loading
. Opening the flap while transfer steps are
being performed results in an immediate stop of the pipettor and entry in the
event log.
2.2.3
Tip Ejection Station, Pipettor Wash Station,
Pipetting Station
Tip ejection station and waste bag
(see figure 2-2: position 3a)
The opening serves as ejection station for disposable tips. The ejected tip is trans-
ported into the waste bag via a slide which is attached to the front side of the instru-
ment. The ejection station is closed by a cover plate which can be pulled off by hand.
With the cover plate removed, the waste bag can be taken out of the holding device
and replaced
(Disposal see chapter 5.3 on page 5-7)
.
Pipettor wash station
The pipettor wash station is located next to the tip ejection station.
T:
Sample rack for 20 patient samples (occupies one track).
R:
Reagent rack for 18 bottles for reagents (occupies 1 tracks).
5:
Reagent rack for 5 large bottles for reagents and 8 small
bottles for reagents (occupies 3 tracks).
Содержание Elisys Quattro
Страница 1: ...HUMAN ELISYS QUATTRO User Manual Elisys Quattro User Manual Cat No 16300 1...
Страница 6: ...TOC 4 ELISYS QUATTRO User Manual HUMAN...
Страница 26: ...2 12 ELISYS QUATTRO User Manual HUMAN System Basics...
Страница 66: ...5 8 ELISYS QUATTRO User Manual HUMAN Loading Unloading Disposal...
Страница 69: ...Service and Maintenance ELISYS QUATTRO User Manual HUMAN 6 3 Pipettor...
Страница 87: ...Technical Data ELISYS QUATTRO User Manual HUMAN 8 5...