5-6
ELISYS QUATTRO
User Manual HUMAN
Loading - Unloading - Disposal
5.2
Unloading
5.2.1
Remove Dilution Tubes / Take Tip Racks off
1.
Pull very left drawer out. The LED DWR must light up green.
2.
Take dilution tube(s) and/or empty tip racks out of holding devices.
Partially used tip racks must remain in the original position. The information
regarding the number and position of the available tips is stored in the
Phoe-
nix
software and this information is used for the subsequent run.
3.
Push drawer in again (audible click).
5.2.2
Remove Racks with Patient Samples and
Reagents
1.
As soon as a LED starts flashing the respective rack can be removed.
Never remove a rack while still in use! Remove it only when the related LED
is flashing! If a sample rack is removed which is still in use, all samples not
already pipetted out of this rack will be flagged with
Removed
and not cal-
culated.
2.
Open flap
.
3.
Pull racks out.
4.
After removal close flap again (audible click).
5.2.3
Remove Test Plate
1.
Open flap. The plate transport unit with the test plate is already in a position
directly next to the plate compartment.
2.
Remove plate with holding frame and take plate off the holding frame.
3.
Close flap again (audible click).
4.
Process plate further manually or dispose off plate.
Содержание Elisys Quattro
Страница 1: ...HUMAN ELISYS QUATTRO User Manual Elisys Quattro User Manual Cat No 16300 1...
Страница 6: ...TOC 4 ELISYS QUATTRO User Manual HUMAN...
Страница 26: ...2 12 ELISYS QUATTRO User Manual HUMAN System Basics...
Страница 66: ...5 8 ELISYS QUATTRO User Manual HUMAN Loading Unloading Disposal...
Страница 69: ...Service and Maintenance ELISYS QUATTRO User Manual HUMAN 6 3 Pipettor...
Страница 87: ...Technical Data ELISYS QUATTRO User Manual HUMAN 8 5...