Spare parts
ENG
Trouble shooting
ENG
d. Check that the actuator cable is connected
properly to control box
e. Please contact Human Care
3. Charging function does not work
a. Check that the charger cable is properly
connected to the lift and mains
b. Check that the battery is properly connected
c. Check that the emergency stop button hasn’t
been pressed
d. If you have an external charger, check if
battery can be charged using the external
charger
e. Please contact Human Care
Spare parts: Part. no FL180
Part. no FL320
Battery 70975
71234
Hand control 70974
71231
Hanger bar L 70977
70977
Hanger bar M 70978
70978
Hanger bar S 70979
70979
Power cable (EU) 70980
70980
Power cable (UK) 70981
70981
Power cable (US) 70982
70982
Power cable (AUS) 70983
70983
Hanger bar 4-point
71366
4. The patient is stuck in a elevated position
a. Check Battery capacity
b. Check that emergency stop hasn’t been
activated
c. Check that hand control cable is properly
connected
d. Use electrical emergency lowering at the
control box by pushing a pen into the
emergency lowering hole
e. Use mechanical emergency lowering at the
actuator
f. Please contact Human care