background image

User Manual

Extension Kit, Floorline LTC

90709

Содержание 90709

Страница 1: ...User Manual Extension Kit Floorline LTC 90709...

Страница 2: ...NGLISH User Manual 3 Compliances and Standards 4 Technical Specifications 5 Responsibility of users 5 Safety Advice 6 Intended Use 7 Assembly Disassembly Instructions 8 Operation Instructions 12 Maint...

Страница 3: ...handy for future use Product modification is not permitted Before use always check that the product is not damaged or worn Human Care s products are continuously being developed and updated to provide...

Страница 4: ...safe handling of this product in its normal and expected environment You should therefore also regard this User Manual as a practical reference book to be kept near the product and readily available...

Страница 5: ...t to hiring rental or lease arrangements Responsibility for the safe operation of this bed accessory lies with the operator User specialist medical staff nurses doctors attendants and care staff are p...

Страница 6: ...t used properly Pay attention to your obligations in order to ensure the permanently safe operation of this product to minimise all risks to the patient user and or third parties Regular preventative...

Страница 7: ...rom Human Care with the following article numbers It is important that only Human Care accessories are fitted to Human Care beds as any incompatible accessories can create hazards Product Name Article...

Страница 8: ...ntion Kit Floorline LTC Assembly Disassembly Instructions Content Number Description Qty 1 Extension Brackets 2 2 Extension Mattress Platform Panel 1 3 Extension Panel Platform Bolts Nuts 4 Extension...

Страница 9: ...with this User Manual and all facets of bed assembly and operation Fitting the extension kit 1 Confirm the kit s contents 2 Clear sufficient space to conduct installation and remove the mattress and a...

Страница 10: ...platorm 16 Now remove the two Foot End Bracket Stepped Pivot Bolts NOTE Keep all nuts bolts and washers safe 17 Carefully lift and move Foot End Pedestal away from the bed frame to a safe place away f...

Страница 11: ...coming loose 30 Reconnect the actuator cable into the mini fit connector securing retaining clip 31 Reconnect the mains power 32 Re check all bed functions with the handset 33 The bed may now be put b...

Страница 12: ...rigger at any time Check that all cables and cords are clear of obstructions that may interfere with bed movement Bed movement may cut cords and cables if they are caught in the movement Before using...

Страница 13: ...Damage Tighten Secure Clean Free Painted steel components Intact Clean Securely Fitted Labels Serial Number Service User Service Manuals In Place Readable PERFORMANCE CHECK Accessories Confirm Full Op...

Страница 14: ...34 Fax 1 613 723 1058 Email info ca humancaregroup com HUMAN CARE UNITED STATES 8006 Cameron Road Suite K Austin TX 78754 USA Phone 1 512 476 7199 Fax 1 512 476 7190 Email info us humancaregroup com H...