Huberg Laser One Скачать руководство пользователя страница 19

 

 

Page 

19

 of 

30

 

 

 

Q.E.D. ENVIRONMENTAL SYSTEMS LIMITED - 

Cyan Park - Unit 3, Jimmy Hill Way, Coventry. CV2 4QP. 

UNITED KINGDOM 

 Tel: +44 (0)333 800 0088 

 E-Mail: 

[email protected]

  

 Web: 

www.qedenv.com

 

REGISTERED IN ENGLAND AND WALES NO. 1898734 

 

OMLASERONE Iss. 01

 

 

Alarms and Error Information 

The table below gives the different alarm conditions or information about errors.  

Displayed Alarm or Error 

Error 

Conditions 

Acoustic alarm and the LED’s 
flash  

GAS alarm 

Concentration measured above the alarm threshold. 

The display shows the message 
ERROR PUMP, LED’s flash and 
the pump icon is crossed 

Pump stopped 

The pump is stopped possibly due to the ingress of 
moisture or a high quantity of dust. -  

Flashing battery icon 

 

Battery level low 

The lowest level of the charge capacity has been reached. 
There is approximately 30 minutes of use left (at +20°C) 

 LOW BATTERY message 
appears for a short duration 

Battery level insufficient   The instrument is not able to work and switches off. 

The display shows: no com from 
laser          

Laser fails to 
communicate properly 

Turn the instrument off and on again. If the error persists, 
contact [email protected] 

The display shows: Laser sensor 
error 130 

Laser fails to stabilise 
properly 

Turn the instrument off and on again. If the error persists, 
contact [email protected] 

The display shows: Laser sensor 
error 135 

Laser fails to stabilize 
properly 

Turn the instrument off and on again. If the error persists, 
contact [email protected] 

The display shows: Laser sensor 
error 150 

Laser fails to stabilize 
properly 

Turn the instrument off and on again. If the error persists, 
contact [email protected] 

The display shows: Laser sensor 
error 142 

Laser fails to stabilize 
properly 

Turn the instrument off and on again. If the error persists, 
contact [email protected] 

The display shows: Laser sensor 
error 144 

Laser fails to stabilize 
properly 

Turn the instrument off and on again. If the error persists, 
contact [email protected] 

Resetting the alarms:  
The table below indicates if it is possible or not to acknowledge and reset the alarm or information 

of default when the device is in operation.  

Alarm information 

Possibility of Reset  

Threshold of concentration 

Sound alarm:  

YES  

Flashing red Led:  

NO 

Pump stopped 

Display indicator:  

NO 

Battery level low 

Display indicator:  

NO 

 

This section outlines various warning and error messages which the operator may receive during general 

operation of the instrument. For further assistance please contact Technical Support at QED on +44(0)333 
800 0088 or email 

[email protected]

 

 

 

Содержание Laser One

Страница 1: ...OPERATING MANUAL...

Страница 2: ...4 2 Restarting the pump operation 13 6 5 SAVING LEAKS 13 7 SETTINGS 14 7 1 MEASUREMENT RANGE 14 7 1 1 Methane measurement 14 7 1 2 Measure in absolute mode ABS and relative mode REL 14 7 2 SETTING THE...

Страница 3: ...Type of equipment Laser One Mains Gas and Water Leak Detection SEM 5000 Landfill Gas Surface Emissions Monitor The Laser One and SEM 5000 described above is in conformity with the relevant Union harmo...

Страница 4: ...100189 Use only this device to charge QED LASER ONE and SEM5000 Maximum voltage Um 15V The maximum voltage that can be applied to the external adapter QED CCLO 100189 to charge QED LASER ONE and SEM5...

Страница 5: ...s T3 for an ambient temperature from 25 C to 50 C The apparatus is category 2 and may be used in areas 1 and 2 The respect of essential safety requirements defined in annex II of the 2014 34 UE direct...

Страница 6: ...easurement of Methane concentrations in sampling mode or in line option The single dial function selector makes the LASER ONE easy to use and the large back lit LCD display provides easy to read menus...

Страница 7: ...edenv com REGISTERED IN ENGLAND AND WALES NO 1898734 OMLASERONE Iss 01 4 USER INTERFACE Overview of the instrument Description 1 ON OFF button Backlight 2 Function button 3 Buzzer 4 Jogdial L LED Alar...

Страница 8: ...atus YES C GPS status NO D Battery status YES E Measurement of Methane YES E1 Type of view ABS absolute REL relative YES E2 Unit of Measurement PPM parts per million VOL volume YES F Bluetooth enabled...

Страница 9: ...1 Navigation System menu Flow Diagram of the jogdial menu Push the jogdial and the menu will appear in the display window E Rotate the jogdial to scroll through the available menu options Once you hav...

Страница 10: ...libration date Current date time and next calibration If there is data stored in the instrument you will see the message DATA TO DOWNLOAD In order to download the data the ProSoft software must be use...

Страница 11: ...lection The GPS icon will change status from Off to Active The percentage of memory used will also be displayed in the bottom left hand corner Initially only the percent symbol will be displayed Once...

Страница 12: ...The use of accessories such as the probes can slightly reduce this flow It is advised to only use the parts and accessories supplied by QED Environmental Systems Limited The sample inlet port is suit...

Страница 13: ...show four horizontal lines Warning The ingress of water or impurities may cause the malfunction of the pneumatic circuit and in some cases the damage to the sensors Warning If the sampling is interru...

Страница 14: ...easurement of the ambient concentrations which will include the background level of Methane present in the atmosphere this is typically between 1 0 and 3 0 ppm It may be desirable for the user to remo...

Страница 15: ...change the ppm threshold by rotating the jogdial to increase or decrease the value To confirm the threshold level push the jogdial Acoustic and visual alarm When the instrument is powered on or if a...

Страница 16: ...rs a red led will be on during the charge cycle which when complete will change to green to indicate a full charge The battery level indicator on the LASER ONE consists of three segments the last of w...

Страница 17: ...E Mail sales qedenv co uk Web www qedenv com REGISTERED IN ENGLAND AND WALES NO 1898734 OMLASERONE Iss 01 8 MAINTENANCE Replacing the Battery Pack Warning Replace the battery pack only in a non hazard...

Страница 18: ...ow regulator 1l min Service and Calibration The LASER ONE should be regularly serviced QED recommends that the instrument is calibrated on an annual basis NOTE The next calibration date can be seen du...

Страница 19: ...ls to stabilise properly Turn the instrument off and on again If the error persists contact technical qedenv co uk The display shows Laser sensor error 135 Laser fails to stabilize properly Turn the i...

Страница 20: ...y QED under warranty the replaced parts must be returned to QED If not returned QED reserve the right to charge for the replacement part If no fault is found an investigation charge may apply QED s Te...

Страница 21: ...equipment has not been maintained in accordance with the operating manual Service Warranty QED offer a three month warranty period following a QED service to cover any defects that have arisen becaus...

Страница 22: ...spects are understood before completing any survey missions Any guarantee may be compromised if the LASER ONE instrument is not handled correctly or damaged as a result of improper use Note QED Enviro...

Страница 23: ...wk tablet for use with Laser One Panasonic carry case FZG1 Power supply Laser one ATEX Charger CCLO for the LASER ONE ATEX Car cable power supply LASER ONE ATEX 900011 100060 100157 507103 402004 Suct...

Страница 24: ...their application Laser radiation The instrument LASERONE contains an invisible laser source The instrument is classified Class 1 according European standards Do not open the device Other precautions...

Страница 25: ...duration 4h30min Battery life 10 hours at 20 C with backlight activated 8 hours at extreme temperatures with backlight activated Case Carbon reinforced polyamide with fiberglass Dimensions L x w x h...

Страница 26: ...sed of through the normal municipal waste stream but through a registered recycling scheme The Waste Electrical and Electronic Equipment directive WEEE makes producers responsible from July 1st 2007 i...

Страница 27: ...nect the equipment into an outlet on a circuit different from that to which the receiver is connected Consult the dealer or an experienced radio TV technician for help Canada The device contains trans...

Страница 28: ...8 E Mail sales qedenv co uk Web www qedenv com REGISTERED IN ENGLAND AND WALES NO 1898734 OMLASERONE Iss 01 Europe The Bluetooth module has been tested to R TTE Directive 1999 5 EC Essential Requireme...

Страница 29: ...NVIRONMENTAL SYSTEMS LIMITED Cyan Park Unit 3 Jimmy Hill Way Coventry CV2 4QP UNITED KINGDOM Tel 44 0 333 800 0088 E Mail sales qedenv co uk Web www qedenv com REGISTERED IN ENGLAND AND WALES NO 18987...

Страница 30: ...TERED IN ENGLAND AND WALES NO 1898734 OMLASERONE Iss 01 www huberg com Contacts SALES ENQUIRIES sales qedenv co uk SERVICE CALIBRATION ENQUIRIES service qedenv co uk TECHNICAL SUPPORT technical qedenv...

Отзывы: