OMLASERONE Iss. 01

 

 

Saving leaks 

To save a leak into the memory of the instrument, push the jogdial and rotate it until the Menu 

“Save Leak” appears. Push the jogdial to save the leak. 

 

This operation is possible only if the GPS is on. 

 

 

Delete data 

To delete the data stored in the instrument, press the jogdial and 

rotate  it  until  the  menu  displays  DELETE  MEMORY.  Press  the 

jogdial to confirm the selection. 

Please Note: There is no secondary confirmation that you want to 

delete data. 

 

 

 

 

Description of the pneumatic circuits 

The LASER ONE samples using a membrane pump. The typical flow is 0.8 l/m and the minimum flow 

is 0.7 l/m. 

The use of accessories, such as the probes can slightly reduce this flow. It is advised to only use the 

parts and accessories supplied by QED Environmental Systems Limited. 

The sample inlet port is suitable for a tube with the dimensions ID2mm X OD4 mm.  

An  internal  hydrophobic  filter  protects  the  device  against  any  ingress  from  dust  or  water.  An 

additional external filter is also recommended which can help protect the internal filter. 

 

 

Warning

 

The Hydrophobic filter is not design to filter anything other than water 
or moisture; chemicals may damage the efficiency of the filter. 

 
Please be vigilant when completing a survey and avoid sampling dust 

and moisture, as if the filter is too saturated it, may result in serious 
damage to the internal sensors.  

 

 

 

Malfunction of the pneumatic circuit / Pump OFF 

 

The  instrument  detects  when  the  pump  flow  is  within  normal 
operating parameters.  
If the flow reduces below certain limits, the pump will stop, the 

LEDs will light up, the pump icon will have a ‘X’ overlay and the 

message "PUMP ERROR" will be displayed. 
 
 

 

Содержание Laser One

Страница 1: ...OPERATING MANUAL...

Страница 2: ...4 2 Restarting the pump operation 13 6 5 SAVING LEAKS 13 7 SETTINGS 14 7 1 MEASUREMENT RANGE 14 7 1 1 Methane measurement 14 7 1 2 Measure in absolute mode ABS and relative mode REL 14 7 2 SETTING THE...

Страница 3: ...Type of equipment Laser One Mains Gas and Water Leak Detection SEM 5000 Landfill Gas Surface Emissions Monitor The Laser One and SEM 5000 described above is in conformity with the relevant Union harmo...

Страница 4: ...100189 Use only this device to charge QED LASER ONE and SEM5000 Maximum voltage Um 15V The maximum voltage that can be applied to the external adapter QED CCLO 100189 to charge QED LASER ONE and SEM5...

Страница 5: ...s T3 for an ambient temperature from 25 C to 50 C The apparatus is category 2 and may be used in areas 1 and 2 The respect of essential safety requirements defined in annex II of the 2014 34 UE direct...

Страница 6: ...easurement of Methane concentrations in sampling mode or in line option The single dial function selector makes the LASER ONE easy to use and the large back lit LCD display provides easy to read menus...

Страница 7: ...edenv com REGISTERED IN ENGLAND AND WALES NO 1898734 OMLASERONE Iss 01 4 USER INTERFACE Overview of the instrument Description 1 ON OFF button Backlight 2 Function button 3 Buzzer 4 Jogdial L LED Alar...

Страница 8: ...atus YES C GPS status NO D Battery status YES E Measurement of Methane YES E1 Type of view ABS absolute REL relative YES E2 Unit of Measurement PPM parts per million VOL volume YES F Bluetooth enabled...

Страница 9: ...1 Navigation System menu Flow Diagram of the jogdial menu Push the jogdial and the menu will appear in the display window E Rotate the jogdial to scroll through the available menu options Once you hav...

Страница 10: ...libration date Current date time and next calibration If there is data stored in the instrument you will see the message DATA TO DOWNLOAD In order to download the data the ProSoft software must be use...

Страница 11: ...lection The GPS icon will change status from Off to Active The percentage of memory used will also be displayed in the bottom left hand corner Initially only the percent symbol will be displayed Once...

Страница 12: ...The use of accessories such as the probes can slightly reduce this flow It is advised to only use the parts and accessories supplied by QED Environmental Systems Limited The sample inlet port is suit...

Страница 13: ...show four horizontal lines Warning The ingress of water or impurities may cause the malfunction of the pneumatic circuit and in some cases the damage to the sensors Warning If the sampling is interru...

Страница 14: ...easurement of the ambient concentrations which will include the background level of Methane present in the atmosphere this is typically between 1 0 and 3 0 ppm It may be desirable for the user to remo...

Страница 15: ...change the ppm threshold by rotating the jogdial to increase or decrease the value To confirm the threshold level push the jogdial Acoustic and visual alarm When the instrument is powered on or if a...

Страница 16: ...rs a red led will be on during the charge cycle which when complete will change to green to indicate a full charge The battery level indicator on the LASER ONE consists of three segments the last of w...

Страница 17: ...E Mail sales qedenv co uk Web www qedenv com REGISTERED IN ENGLAND AND WALES NO 1898734 OMLASERONE Iss 01 8 MAINTENANCE Replacing the Battery Pack Warning Replace the battery pack only in a non hazard...

Страница 18: ...ow regulator 1l min Service and Calibration The LASER ONE should be regularly serviced QED recommends that the instrument is calibrated on an annual basis NOTE The next calibration date can be seen du...

Страница 19: ...ls to stabilise properly Turn the instrument off and on again If the error persists contact technical qedenv co uk The display shows Laser sensor error 135 Laser fails to stabilize properly Turn the i...

Страница 20: ...y QED under warranty the replaced parts must be returned to QED If not returned QED reserve the right to charge for the replacement part If no fault is found an investigation charge may apply QED s Te...

Страница 21: ...equipment has not been maintained in accordance with the operating manual Service Warranty QED offer a three month warranty period following a QED service to cover any defects that have arisen becaus...

Страница 22: ...spects are understood before completing any survey missions Any guarantee may be compromised if the LASER ONE instrument is not handled correctly or damaged as a result of improper use Note QED Enviro...

Страница 23: ...wk tablet for use with Laser One Panasonic carry case FZG1 Power supply Laser one ATEX Charger CCLO for the LASER ONE ATEX Car cable power supply LASER ONE ATEX 900011 100060 100157 507103 402004 Suct...

Страница 24: ...their application Laser radiation The instrument LASERONE contains an invisible laser source The instrument is classified Class 1 according European standards Do not open the device Other precautions...

Страница 25: ...duration 4h30min Battery life 10 hours at 20 C with backlight activated 8 hours at extreme temperatures with backlight activated Case Carbon reinforced polyamide with fiberglass Dimensions L x w x h...

Страница 26: ...sed of through the normal municipal waste stream but through a registered recycling scheme The Waste Electrical and Electronic Equipment directive WEEE makes producers responsible from July 1st 2007 i...

Страница 27: ...nect the equipment into an outlet on a circuit different from that to which the receiver is connected Consult the dealer or an experienced radio TV technician for help Canada The device contains trans...

Страница 28: ...8 E Mail sales qedenv co uk Web www qedenv com REGISTERED IN ENGLAND AND WALES NO 1898734 OMLASERONE Iss 01 Europe The Bluetooth module has been tested to R TTE Directive 1999 5 EC Essential Requireme...

Страница 29: ...NVIRONMENTAL SYSTEMS LIMITED Cyan Park Unit 3 Jimmy Hill Way Coventry CV2 4QP UNITED KINGDOM Tel 44 0 333 800 0088 E Mail sales qedenv co uk Web www qedenv com REGISTERED IN ENGLAND AND WALES NO 18987...

Страница 30: ...TERED IN ENGLAND AND WALES NO 1898734 OMLASERONE Iss 01 www huberg com Contacts SALES ENQUIRIES sales qedenv co uk SERVICE CALIBRATION ENQUIRIES service qedenv co uk TECHNICAL SUPPORT technical qedenv...

Отзывы: