This instrument provides two shutdown modes:
★
timed shutdown: when the timing reaches the set time, the instrument will stop automatically.
★
manual shutdown: click the "shutdown" key to stop the machine. The two
shutdown modes are normal shutdown.
★
there are also two kinds of abnormal shutdown: over-voltage protection shutdown and over-
current protection shutdown.
★
overvoltage protection shutdown
In the process of test, when the output high pressure exceeds the set limit, the instrument will
automatically cut off the output after starting and shutting down the command, and then execute
the data history saving. After shutting down, as shown in Figure 12, the prompt message in the
status bar is "over voltage protection" and execute the data history saving.
Fig. 12 controller prompts over voltage protection
★
over current protection shutdown
In the process of test, when the output current exceeds the set limit, the instrument will
automatically cut off the output after starting and stopping the command, and then