Charging:
a. Insert the adapter to battery charging port. Connect adapter power to AC
power socket.
b. When adapter light is green means finish charging.
Note: Longest charging time is 6-8 hour please don’t charge it over.
If instrument doesn’t reflect anything failed to turn off please open back cover get off
the battery and insert it again then turn on the device.
If touch screen is not smart please calibrate it in “System management”.
When screen display “Insert storage card again” please get off the SD card and
insert it again or replace SD card.
If we don’t operate it for a long while during test the instrument will close LCD or
whole device automatically. Click the touch screen or press any key will wake it up.
We can set the back light and off time ourselves.
5.2 Display
“
Inspect batter clip
”
Firstly ensure battery clips and pole are connected well and testing battery voltage
should be not less than 1.5V.
5.3 Test pin probe bent or can
’
t withdraw
Replace the probe:
a. Clip the probe on sleeve top by pincher.
b. Warning: Don’t clip the sleeve holding probe to avoid damaging pin.
c.
Pull out the pin probe straightly.
d. Clip new probe into sleeve.
e. Push the probe to a soft surface. Such as cardboard until the probe reach the
sleeve top.