8
Fig.3.2.1Specifications setting
New battery specification: Custom setting on battery test specification.
Existing battery specification: Specifications of former test battery.
Fig.3.2.1-1New battery specification setting
Machine room number: 0001-9999.
Technician number: 0001-9999.
Battery group number: 0001-9999.
Battery quantity: 1-500(The continuous test can be made for 500 pcs battery at
most).
B.Brand:
a. Select USER, set the single battery voltage, volume, reference internal
resistance specification according to actual demand.
b. When we select battery manufacturer, input voltage, battery volume setting via
selection orderly the reference internal resistance will display automatically but
also can be input by hand.
Single voltage: Rated voltage of battery.
Battery volume: Battery rated volume.
Reference internal resistance: Refer to battery manufacturer instruction.
Minimum voltage setting. During test if voltage is lower than this level device will
send alarm but won’t affect test.
Test mode:
:
a.
By manual or automatic. If manual test each change on test clip will request us to
click “Start test” for next test.
b.
Automatic test. Every time change on test clip the instrument will test battery