4

light is off

Poor power connection

Check the circuit for short

circuit

The power switch is damaged

Replace the power switch

Turn on the

water bath

stirring switch,

the motor does

not rotate

There is a problem with the

stirring switch

Replace the stirring switch

There is a problem with the

stirring motor

Replace the mixing motor

Open circuit

Check and exclude

The bath

temperature is

not constant

Solid state relay is broken

Replace solid state relay

Damaged heating tube

Replace heating tube

Sensor damaged

Replace the sensor

Temperature control table is

damaged

Replace the temperature

control table

The timer does

not show

Timer open

Check the line

The timer does

not count or

clear

Bad timer contact

Check the connection

Timer malfunction or button

switch is broken

Replace timer or button

switch

VI. Precautions

1. Keep the instrument clean, do not get water or damp into the electronic control part,

and prevent dirt such as oil samples from being spilled on the instrument.

2. The instrument should be well grounded before use to ensure the safety of operators.

3. When the instrument fails, please professional and technical personnel to deal with it,

and do not disassemble it at will.

4. This instrument is guaranteed for one year.

Содержание HZ-1025

Страница 1: ...HZ 1025 Petroleum and synthetic liquid water separability tester Huazheng Electric Manufacturing Baoding Co Ltd...

Страница 2: ...Contents I Overview 1 II Key Function and Feature 1 III Major Technical Indicators 2 IV Assembly and use 2 V Fault analysis and troubleshooting 3 VI Precautions 4 VII Packing List 5...

Страница 3: ...mm s at 40 but the test temperature is 82 1 II Key Function and Feature 1 Temperature control table 2 Chronograph 3 Power switch 4 Mixing switch 5 Sample stirring switch 6 Lighting switch 7 Reset swit...

Страница 4: ...ug and heating tube plug into the corresponding sockets on the rear panel 3 Clamp the test tube containing the sample with a clamp and place it on the upper cover steadily 4 Connect the power supply a...

Страница 5: ...key press The function parameters are shown in the table below Prompt name set range description Initial value RESET 99 99 9 100 100 0 RESET Only used to adjust the static difference of proportional c...

Страница 6: ...ce the sensor Temperature control table is damaged Replace the temperature control table The timer does not show Timer open Check the line The timer does not count or clear Bad timer contact Check the...

Страница 7: ...ies NO Name Specification type Unit Quantity 1 Certificate and manual Pieces 1 2 Host Set 1 3 Measuring cylinder 100ml Pieces 3 4 Test tube fixture Set 3 5 light Pieces 1 6 Test stirring device Set 1...

Отзывы: