background image

1

I. Overview

The HZ-1025 petroleum and synthetic liquid water separability tester is designed and

manufactured in accordance with the national standard GB/T7305 ASTM D1401. This
instrument is suitable for measuring oils with a kinematic viscosity of 28.8

90mm

²

/s at

40

. The test temperature is 54

±

1

. It can also be used to determine oils with a

kinematic viscosity greater than 90mm

²

/s at 40

, but the test temperature is 82

±

1

.

II. Key Function and Feature

1-Temperature control table

2-Chronograph

3-Power switch

4-Mixing switch

5-Sample stirring switch

6-Lighting switch

7-Reset switch

8-lamp holder

9-Sample stirring motor

10-Pole

11-Stirrer

The HZ-1025 petroleum and synthetic liquid water separability tester is mainly composed

of thermostat, control device and heating device.

The thermostat is composed of a thermostat and a stirring motor.

The control device consists of a potentiometer and a control switch.

The heating device is composed of a heating tube.

Содержание HZ-1025

Страница 1: ...HZ 1025 Petroleum and synthetic liquid water separability tester Huazheng Electric Manufacturing Baoding Co Ltd...

Страница 2: ...Contents I Overview 1 II Key Function and Feature 1 III Major Technical Indicators 2 IV Assembly and use 2 V Fault analysis and troubleshooting 3 VI Precautions 4 VII Packing List 5...

Страница 3: ...mm s at 40 but the test temperature is 82 1 II Key Function and Feature 1 Temperature control table 2 Chronograph 3 Power switch 4 Mixing switch 5 Sample stirring switch 6 Lighting switch 7 Reset swit...

Страница 4: ...ug and heating tube plug into the corresponding sockets on the rear panel 3 Clamp the test tube containing the sample with a clamp and place it on the upper cover steadily 4 Connect the power supply a...

Страница 5: ...key press The function parameters are shown in the table below Prompt name set range description Initial value RESET 99 99 9 100 100 0 RESET Only used to adjust the static difference of proportional c...

Страница 6: ...ce the sensor Temperature control table is damaged Replace the temperature control table The timer does not show Timer open Check the line The timer does not count or clear Bad timer contact Check the...

Страница 7: ...ies NO Name Specification type Unit Quantity 1 Certificate and manual Pieces 1 2 Host Set 1 3 Measuring cylinder 100ml Pieces 3 4 Test tube fixture Set 3 5 light Pieces 1 6 Test stirring device Set 1...

Отзывы: